5. MAINTENANCE page 33
5.MAINTENANCE
5.1.OPERATOR MAINTENANCE
The slide tracks on the weight scale platform should be greased monthly. The hinges and door latch do not require lubrication.
5.2. CALIBRATION
The electronic circuitry has proven to be very stable over long periods of time. However, the following calibration should be performed at least annually to ensure reliable operation and accurate measurements. There is a Calibration Checklist (5.5.) at the end of this section. Copies should be made, and filled out each time the system is calibrated.
NOTE:
The high voltage affects the pulse amplitude, so the high voltage must be tested before
the
Calibration equipment required:
DVM with 10MÛ input impedance
Oscilloscope with >20 MHz bandwidth
Test Source
Remove the right side face plate to gain access to the RCS boards. Refer
Model