6. TROUBLESHOOTING page 39

6. TROUBLESHOOTING

This guide is designed so that on-site personnel can service the BM-286A and effect necessary minor repairs. It covers procedures and parts down to the board level. Any other problems should be referred to factory authorized service personnel. Unauthorized repair voids warranty.

6.1. DIAGNOSTIC AIDS

There are several diagnostic tools available to help isolate problems in the unit. Access to the diagnostics is available only to level 0 users with a valid password. The system administrator normally controls the password access. Press Utilities from the operating screen to access the Utilities Menu. From the Utilities Menu, the following diagnostics may be selected:

6.1.1. Scale Reset:

Scale Reset will power cycle the weight scale controller.

6.1.2. Variance:

Variance tests the statistics of the detector counts versus a Gaussian distribution. If periodic noise is being counted the results of the variance test will approach 1.0. Light leaks will result in a variance near 2.0. Refer to section 3.5.1 for details on the variance test.

6.1.3. Detectors, Scale and Disc.:

The Detectors, Scale and Disc. diagnostic provides a convenient way to observe detector counts with a relatively quick update (2 seconds). Alarms and door position are ignored in this mode.

6.1.4. Detector Efficiency Check:

Detector Efficiency Check performs an efficiency calibration on the six plastic scintillation detectors. The results are only displayed, and are not written to disk.

Model BM-286A

Image 42
Contents Waste Monitor Model BM-286A Maintenance Table of ContentsInspection and Installation Theory of OperationConfiguration Tracking Sheet Parts TroubleshootingAppendix Warranty Glossary Formulas Recommended Spare Parts Spare Parts Ordering InformationIntroduction Scope and Purpose of ManualSpecifications Introduction Introduction Inspection Installation Operating Instructions General OverviewPOWER-UP and SELF-TESTS Access Menu Level One AccessPCA-II Normal Operation Operating Instructions Programming the System SET-UP Menu AbcdefghOperating Instructions Operating Instructions Waste Stream Parameters Operating Instructions Operating Instructions Operating Instructions This value is used only when the Alarm Type is set to DPM Utilities Menu Operating Instructions Reports Initial Installation Checklist Operating Instructions BM-286A Waste Stream SET-UP Worksheet Unit Operation Theory of Operation Theory of Operation Theory of Operation Theory of Operation Operator Maintenance Maintenance Maintenance Settings for Plastic Detectors LLD ULD Weight Calibration Maintenance Diagnostic Aids Fault Isolation Computers Longs Peak Court LONGMONT, Colorado Appendix B. Glossary Appendix B. Glossary Activity from Counts Reliable Detectable Activity RDA Formula Variance Configuration Tracking Sheet System ModificationsLongs Peak Court LONGMONT,COLORADO Drawings