Page 10

Rev 05

Positioning

the Patient.

When lifting the anterior coil housing, always grasp the coil by the apron and the superior end of the housing. Never lift the coil housing by the viewing window or mirror.

Adjust the patient so their shoulders are snug against the curved arch of the posterior coil housing.

Firmly hold the anterior coil housing and carefully place it on the posterior coil housing. As you lower the anterior coil housing, center and seat the RF connector pins while the flat mating surfaces meet. Secure the latches on each side of the housing.

Slide the viewing mirror if necessary for patient viewing. Use the window on the anterior coil housing to ensure the patient is centered properly. Using posititoning beams, center on the anatomy to be imaged.

10

Image 10
Contents Rev Licenses and Trademarks Introduction CompatibilityHNC Neurovascular Array Coil GrasphereTable of Contents Precautions Patient SafetyTraining Quality Assurance Indications Contra- indications Emergency Procedures Technical Considerations Array Coil Positioning the HNC-63-INT Coil on the Patient TablePositioning Patient Synergy MultiConnect SMC Connecting Cable Left Side Patient Interface Control Unit Picu Right Side Patient Interface Control Unit PicuQuality Assurance MTC Rev Save raw data Yes Rev SNR Calulations Example SNR Calulations SNR Calulations Example SNR Calulations Scanning Set Up Head and Neck Imaging Selecting the Active CoilScanning Step by Step Scan Protocols Cleaning StorageRev