Reconstruct SNR CHECK image for individual channels using the instructions provided below.

1.Before running the SNR CHECK scan, be sure to have the "Save Raw Data" option set to "Yes" under [Postproc].

2.Do not delete the current scan from the [Scanlist].

3.Under the [Scanlist] icon, select the [Add Rec] icon, then the [Del. Recon] icon.

4.Left-mouse select on the scan to perform the reconstructions, then select [Proceed].

5.Change the "Synergy selection" option from "0" to "2" for reconstruction of the 2nd channel, then select [Proceed].

6.Select [Options], and change the name of the reconstruction to "2".

7.Repeat steps 3-6 for each channel (channels 3,4,5), and change the "Synergy selection" option to the next channel, then rename the scans appropriately for each channel selected (i.e. 2,3,4,5)

8.From the [Scan List] icon, delete all scans except those renamed for each element.

9.Select [Start Scan] to start the reconstruction of the images.

Note: Example of reconstructed images are shown below.

18

Image 18
Contents Rev Licenses and Trademarks Introduction CompatibilityHNC Neurovascular Array Coil GrasphereTable of Contents Patient Safety PrecautionsTraining Quality Assurance Indications Contra- indications Emergency Procedures Technical Considerations Array Coil Positioning the HNC-63-INT Coil on the Patient TablePositioning Patient Synergy MultiConnect SMC Connecting Cable Left Side Patient Interface Control Unit Picu Right Side Patient Interface Control Unit PicuQuality Assurance MTC Rev Save raw data Yes Rev SNR Calulations Example SNR Calulations SNR Calulations Example SNR Calulations Scanning Set Up Head and Neck Imaging Selecting the Active CoilScanning Step by Step Scan Protocols Cleaning StorageRev