SMC Networks 500082, HNC-63-INT manual Save raw data Yes

Page 17

Rev 05

Run an SNR CHECK scan using the parameters listed below. TIP: Save this SNR CHECK scan for future QA scans.

Position slices for the SNR CHECK as shown on page 15.

***IMPORTANT: Be certain to SAVE RAW DATA under Post Processing.***

Geometry

 

CoilSelection

3IB-MRIDevic

-channelCombination

2345

-connection

d

Homogeneitycorrection none

FOV(mm)

400.00

RFOV(%)

100.00

Foldoversuppression

no

Matrixscan

256

-reconstruction

256

Scanpercentage(%)

100.00

Stacks

1

-type

parallel

-slices

3

-slicethickness(mm)

5.00

-slice gap

userdefined

- - gap (mm)

18.00

-sliceorientation

sagittal

-foldover direction

AP

Slicescanorder

default

PlanAlign

no

RESTslabs

0

Patientposition

headfirst

-orientation

supine

Contrast

 

Scanmode

MS

-technique

FFE

Contrastenhancement

no

FastImagingmode

none

Echoes

1

-partialecho

yes

TE

userdefined

-(ms)

15.00

Flipangle(deg)

30.00

TR

userdefined

-(ms)

200.00

HalfScan

no

Water fat shift

userdefined

-(pixels)

2.000

Shim

no

SPIR

no

MTC

no

SAR mode

default

Gradient mode

regular

Motion

 

Cardiac synchronisation

no

Respiratory compensation

no

Flow compensation

yes

NSA

2

SMART

no

Dyn/ang

 

Angio

no

Quantitative flow

no

Manual start

no

Dynamic study

no

Postproc

 

Preparation phases

full

MIP/MPR

no

Images:

O:M no no no

Autoview image

M

Reference tissue

Grey matter

Preset window contrast

soft

Reconstruction mode

immediate

***Save raw data***

yes

Push to workstation

no

Hardcopy protocol

no

Ringing filter

no

Silent mode

no

Offc/ang

 

Stacks

1

Stack Offc.AP (P=+mm)

-20.00

RL (L=+mm)

0.00

FH (H=+mm)

0.00

Ang. AP (deg)

0.00

RL (deg)

-0.00

FH (deg)

-0.00

17

Image 17
Contents Rev Licenses and Trademarks Compatibility IntroductionGrasphere HNC Neurovascular Array CoilTable of Contents Training Quality Assurance Indications Contra- indications Patient SafetyPrecautions Emergency Procedures Technical Considerations Positioning the HNC-63-INT Coil on the Patient Table Array CoilPositioning Patient Synergy MultiConnect SMC Connecting Cable Right Side Patient Interface Control Unit Picu Left Side Patient Interface Control Unit PicuQuality Assurance MTC Rev Save raw data Yes Rev SNR Calulations Example SNR Calulations SNR Calulations Example SNR Calulations Head and Neck Imaging Selecting the Active Coil Scanning Set UpScanning Step by Step Scan Protocols Storage CleaningRev