Technical Considerations

The coil and accessories require special conditions regarding electromagnetic compatibility. The coil must be installed and used in a shielded scan room provided with the MR magnet and system. The user must ensure that the scan room door is closed during system use. Failure to do so may cause reciprocal interference with portable and mobile RF communications equipment, affecting the performance of the MR coil and/ or such equipment.

The coil should only be used with the accessories specified in the operator’s manual.

The use of accessories other than those specified in the operator’s manual may result in decreased ESD immunity of the coil or MR system, causing damage to the coil and/or system.

The equipment should not be used with other coils or equipment present in the MR scanner except as specified in the Operator’s Manual.

Tampering with the cable pins and connector may damage the connector and affect coil or system performance. Please verify that connector and pins are not damaged before use.

8

Image 8
Contents Rev Licenses and Trademarks Introduction CompatibilityHNC Neurovascular Array Coil GrasphereTable of Contents Training Quality Assurance Indications Contra- indications Patient SafetyPrecautions Emergency Procedures Technical Considerations Array Coil Positioning the HNC-63-INT Coil on the Patient TablePositioning Patient Synergy MultiConnect SMC Connecting Cable Left Side Patient Interface Control Unit Picu Right Side Patient Interface Control Unit PicuQuality Assurance MTC Rev Save raw data Yes Rev SNR Calulations Example SNR Calulations SNR Calulations Example SNR Calulations Scanning Set Up Head and Neck Imaging Selecting the Active CoilScanning Step by Step Scan Protocols Cleaning StorageRev