Rev 05
Chapter 4 - Scanning Set Up
Head and Neck Imaging
Selecting the Active Coil
One of the advantages of using the Invivo
The neurovascular array coil contains both a head coil and a neck coil. The head coil (channel 2), neck coil (channels 3,4,5) or both head and neck coils (channels 2,3,4,5) can be active at any given time. To select the active coil(s), pick the proper channel(s) from the channel combination (ch. combination) menu on your Intera Operator's Console.
Field of View and Coverage
Survey Image
IMPORTANT! Using Autoshim
EXAMPLES:
HEAD only: Select | Ch. combination 2 |
NECK only: Select | Ch. combination 3,4,5 |
HEAD & NECK: Select | Ch. combination 2,3,4,5 |
For head or neck studies a FOV of 200 mm. is suggested, depending upon head size. The
The Intera body coil may be used at any time while the
mm)body coil SURVEY to be performed, which is helpful in determining the
AUTOSHIM is a feature of Intera software to improve image quality by improving the magnetic field homogeneity within the FOV selected. The improvement in image quality is often dramatic when the selected FOV is far off center, and when acquiring SPIR images.
23