SMC Networks 500082, HNC-63-INT manual Mtc

Page 15

Rev 05

Run a SURVEY scan using the parameters listed below.

TIP: Save this SURVEY scan for future QA scans.

An example of the SURVEY images is on page 15.

Geometry

 

SPIR

no

CoilSelection

3IB-MRIDevic

TFEprepulse

invert

-channelCombination

2345

-sliceselection

no

-connection

d

-delay

shortest

Multi Coil

no

MTC

no

Homogeneitycorrection none

SARmode

default

FOV(mm)

400.00

Gradient mode

regular

RFOV(%)

100.00

 

 

Foldoversuppression

no

Motion

 

Matrixscan

256

 

 

-reconstruction

256

Cardiac synchronisation

no

Scanpercentage(%)

50.00

Respiratory compensation

no

Stacks

3

Flow compensation

no

-current

A

NSA

1

-type

parallel

 

 

-slices

3

Dyn/ang

 

-slicethickness(mm)

10.00

Angio

no

-slice gap

userdefined

Quantitative flow

no

- - gap (mm)

10.00

Manual start

no

-sliceorientation

sagittal

Dynamic study

no

-foldover direction

AP

 

 

Slicescanorder

default

Postproc

 

Stackdisplayorder

no

Preparation phases

auto

PlanAlign

no

MIP/MPR

no

RESTslabs

0

Images:

O:M no no no

Patientposition

headfirst

Autoview image

M

-orientation

supine

Reference tissue

White matter

 

 

Preset window contrast

soft

Contrast

 

Reconstruction mode

immediate

Scanmode

M2D

Save raw data

no

-technique

FFE

Push to workstation

no

Contrastenhancement

T1

Hardcopy protocol

no

FastImagingmode

TFE

Ringing filter

no

-shotmode

multi-shot

Silent mode

no

TFEfactor

42

 

 

-shot interval

shortest

Offc/ang

no

-profile order

linear

Stacks

3

Echoes

1

-current

A

-partialecho

no

Stack Offc.AP (P=+mm)

-20.00

TE

shortest

RL (L=+mm)

0.00

Flipangle(deg)

20.00

FH (H=+mm)

0.00

TR

userdefined

Ang. AP (deg)

0.00

-(ms)

15.00

RL (deg)

-0.00

HalfScan

no

FH (deg)

-0.00

Water fat shift

maximum

 

 

Shim

no

 

 

15

Image 15
Contents Rev Licenses and Trademarks Compatibility IntroductionGrasphere HNC Neurovascular Array CoilTable of Contents Patient Safety PrecautionsTraining Quality Assurance Indications Contra- indications Emergency Procedures Technical Considerations Positioning the HNC-63-INT Coil on the Patient Table Array CoilPositioning Patient Synergy MultiConnect SMC Connecting Cable Right Side Patient Interface Control Unit Picu Left Side Patient Interface Control Unit PicuQuality Assurance MTC Rev Save raw data Yes Rev SNR Calulations Example SNR Calulations SNR Calulations Example SNR Calulations Head and Neck Imaging Selecting the Active Coil Scanning Set UpScanning Step by Step Scan Protocols Storage CleaningRev