Omega OMB-DAQBOARD-500 manual Configure Boards, Accessing the DaqBoard/500 Properties Tab

Page 11
Step 3 – Configure Boards

Step 3 Configure Boards

DaqBoard/500 Series Boards have no jumpers or switches to set. Configuration is performed entirely through software. Refer to the following figure and steps to complete the configuration. The numbers in the figure correspond to the numbered steps.

1.Open the “Start” menu from the Windows desktop.

2.Select “Settings.”

3.Select “Control Panel.”

4.Double-click “Daq Configuration.” This opens the Daq Configuration window.

5.Double-click on the Device Inventory’s DaqBoard/500 Series icon. In the figure above the DaqBoard/500 appears as “DaqBoard500_0.”

The DaqBoard’s Properties tab will appear (following figure).

Note: If the DaqBoard icon is not present, skip to the upcoming section, Using ‘Add Device.’

Accessing the DaqBoard/500 Properties Tab

6.Enter a “Device Name” in the text box, or use the default, e.g., DaqBoard500_0. The Name is for identifying the specific DaqBoard, but actually refers to the PCI slot.

7.Verify that the “Device Type” shows the correct board, i.e., “DaqBoard/500 or DaqBoard/505.” Other devices, if available can be viewed via the pull-down list.

8.Confirm that the DaqBoard’s text box shows a Bus #, Slot #, and Serial Number.

If this text box is empty, use its pull-down list and select the serial number that matches the one for your board.

1138-0940, rev 1.1

947994

DaqBoard/500 Series Installation Guide IG-5

Image 11
Contents Shop online at 16-Bit, 200-KHz PCI Data Acquisition BoardsUser’s Guide OMB-DAQBOARD-500 SeriesFor immediate technical or application assistance OMEGAnet Online Service omega.comInternet e-mail info@omega.com Servicing North AmericaWarnings, Cautions, Notes, and Tips Specifications and Calibration⇒ Programs ⇒ Omega DaqX Software 6 - Specifications Glossary Table of ContentsOMB-DAQBOARD-500 Series, Installation Guide 1 - Introduction 3 - ConfigurationThis product requires one of the following Operating Systems Windows Windows XPReference Note OMB-DAQBOARD-500 SeriesInstallation Guide DaqView/500 Authorization CodeReference Notes Step 1 - Install SoftwareIMPORTANT Software must be installed before installing hardware Installation, A Pictorial Overview2. Remove the PC’s cover. Refer to your PC Owner’s Manual as needed Step 2 - Install Boards in available PCI Bus-slotsIMPORTANT Bus Mastering DMA must be Enabled 9. Secure the board by inserting the rear-panel adapter-plate screw Accessing the DaqBoard/500 Properties Tab Step 3 - Configure BoardsTest Hardware Tab Using “Add Device”Step 4 - Test Hardware Basic Information IntroductionDaqBoard/500 Series User’s Manual Block DiagramDaqBoard/500 Series Block Diagram Digital I/O Board FeaturesTimer 0 and Timer Analog I/O947294 Overview Connections and PinoutsPower 68-Pin SCSI Type III PinoutStandard 68-Pin SCSCI Type III, Socket Female Connector with Orb SignalHI + Pin # Signal DefinitionsAnalog Input Channels Differential Channels, Pin ReferenceClocks, Triggers, Counters, and Timers Ground Lines Power Line2-6 Connections & Pinouts TB-100 Terminal Connector OptionScrew Terminals for TB2 Side Screw Terminals for TB1 SideHI + Connecting User WiringExternal Connections Differential Channels Screw Terminal ReferenceSingle-Ended Signal TypesChoosing A/D Input Configuration Single-Ended ConfigurationDifferential Configuration for Grounded Sources Differential DIFFDifferential for Grounded Signal Sources Differential Configuration for Floating Sources Differential for Floating Signal SourcesIf the Input Signal is DC Coupled If the Input Signal is AC CoupledADC Ranges ConfigurationConfiguration through Software Analog Input ConfigurationMicrovolt Resolution DAC RangesDAC OUTPUT DaqBoard/500 DAC Range Resolution Range ConfigurationDrivers for Third-party, Icon-driven Software …… DaqCOM Driver …… Software and Board OperationOut-of-theBoxTM Software Overview …… Out-of-the-BoxTM Software ……4-2 Software and Board Operation Drivers for Third-party, Icon-driven SoftwareDaqCOM Driver LabVIEWProcess Definitions Theory of OperationADC Maximum Clock Rate Clocking the ADCADC Pacer Clocking ADC External Event ClockingADC External Trigger Starting Triggering an ADC AcquisitionADC Software ADC External GateStopping an ADC Acquisition CLOCK ADC Clock and FIFO ErrorsDaqBoard/500 Only Digital AcquisitionDAC Software Update Clocking the DAC4-8 Software and Board Operation Emissions/Immunity Conditions ……5-2 CE-ComplianceCE Standards and Directives Overview ……5-1 CE Standards and Directives …… Safety Conditions ……5-2Emissions/Immunity Conditions Safety Conditionsof calibration SpecificationsGeneral Analog Inputs6-2 Specifications Analog Input Triggering/ClockingAnalog Outputs DaqBoard/500 Only Digital Inputs / OutputsCounter - 1 Counter designated as CNTR1 Physical & Environmentalsingle-ended mode Glossaryreferenced to common Page Page RETURN REQUESTS/INQUIRIES WARRANTY/DISCLAIMERTEMPERATURE 324545A-01Shop online at omega.com OMEGA…Of Course