E: Specifications
Power Information
Power Requirements
Voltage: | 95 | - 250 Volts AC, |
Frequency: | ||
Operating Current: | 300 mA @ 120 V | |
Power Consumption: | 40 | Watts (maximum) |
Fuse Rating: | 1.6A, 250 Volts |
Power Supply Cord
Cord type: | 3 conductors, 1.0 mm2 minimum conductor size |
| (approximately 18 AWG) |
Rated for: | 250 Volts AC, 10 Amps |
Length: | 3.0 meters |
Environmental Limitations |
|
Temperature |
|
Operating range: | 5° to 50° C (41° to 122° F) |
Storage range: | |
Max temp. change/hr: | 20° C (36° F) |
Rapid temperature changes may affect operation. Therefore, do not operate the ETS near heating or cooling devices, large windows, or doors that open to the outdoors.
Altitude
Operating maximum: | 2.4 km (8,000 ft) |
Storage maximum: | 9.1 km (30,000 ft) |
If operating the ETS above 2.4 km (8000 ft.), decrease the operating temperature rating by 1° F for each 1000 ft.
Relative Humidity
Operating range: | 10% to 90% noncondensing, 40% to 60% |
| recommended |
Storage range: | 10% to 90% noncondensing |
ETS Installation Guide | 60 |