_______________________________________________________________________________________APPLICATIONS

Two Identical Negative Voltages With a Common Positive (Refer To Fig. 13)

When the same negative voltage is required at two points in the same circuit and a common positive is needed, perform the following:

1.Select the TRACKING PARALLEL operating mode and set the A/B Metering switch to monitor the “A” supply.

2.Set the desired voltage and maximum current using the “A” VOLTAGE and CURRENT controls.

3.Connect a ground wire between the “A” supply positive terminal and the GND (green) terminal.

4.Turn off the power supply and the equipment to be powered during hook-up.

5.Connect the negative polarity inputs of the circuit to be powered to the negative (black) terminals of the supplies. Connect the common positive input of the circuit to be powered to the “A” supply’s positive (red) or GND (green) terminal.

33

Fig. 13. Typical Hook-Up Using Two Identical Negative Voltages and a Common Positive.

Image 37
Contents Instruction Manual Test Instrument Safety DC Power Supply Table of Contents Introduction Introduction Features Overload Protection Isolated OutputReverse Polarity Protection HOOK-UP CablesB Supplies SpecificationsGeneral Fixed 5V SupplySupply Controls and Indicators Controls and IndicatorsControls and Indicators Constant Current Indicator Line Voltage Select Switches. Combination Fixed 5V Supply Terminals and Indicator Rear Panel ControlsRear panel controls Equipment Precautions Safety PrecautionsOperating Instructions Independent USE of a or B SupplyOperating Instructions Hook-up Operating Instructions Independent operation grounding possibilities sheet 2 Setting Current Limit Setting Current limit Series Tracking Operation Constant voltage/constant current characteristicOperating Instructions Operating Instructions Operating Instructions Parallel Tracking Operation Parallel tracking operation grounding possibilities sheet 1 Parallel tracking operation grounding possibilities sheet 2 Fixed 5 V Power Supply Operation Grounding possibilities for fixed 5 V power supply sheet 1 Grounding possibilities for fixed 5 V power supply sheet 2 Operating Instructions Electronics Servicing ApplicationsElectronics Manufacturing Electronics Design LAB Battery ChargingElectronics Education Split SupplyPage Page Page Page Page Page Fuse Replacement MaintenanceService Information Limited One-Year Warranty Test Instrument Safety Manual DE Instrucciónes Fuente DE Poder DE CD Triple Seguridad DEL Instrumento DE Prueba Advertencia Instrucciones DE Operación Seguridad DEL Instrumento DE PruebaIntroduccion Fuente Fija DE Salida TripleDOS Fuentes DE O a Característica DE Rastreo UnicaProtección Contra Sobreflujo Salida AisladaProtección DE Polaridad Inversa Cables Fuentes a Y B EspecificacionesGeneral Controles E Indicadores Figura 1. Controles e indicadores del panel frontal Controles E Indicadores DE LA Fuente B Indicador Constant CurrentControles DEL Panel Trasero Indicadores Y Controles DE LA Fuente DEFigura 2. Controles del panel trasero USO Independiente DE LA Fuente a O B Precauciones Para SeguridadInstrucciones DE Operación Page Page Figura 4. Operación de voltaje constante típica Operación de corriente constante típica Operación DE Rastreo EN Serie Característica de voltaje constante/corriente constantePage Page Instrucciones DE Operación Operación DE Rastreo EN Paralelo Page Page Operación DE LA Fuente Fija DE Page Page Instrucciones DE Operación Manufactura Electrónica Servicio ElectrónicoEducación EN Electrónica Laboratorio DE Diseño ElectrónicoCarga DE Baterias Fuente DivididaAplicaciones Aplicaciones Aplicaciones Aplicaciones Aplicaciones Aplicaciones Mantenimiento Información de Servicio Garantía Limitada de Un Ano Seguridad DE EL Equipo DE Prueba Declaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA 2005 B&K Precision Corp