Identical Positive and Negative Voltages With a Separate Common (Refer To Fig. 15)

Another typical “split supply” application is when a circuit uses operational amplifiers (op-amps). Typically, identical positive and negative voltages are required to power op-amp circuits. Using both supplies and the series tracking mode of operation, identical positive and negative voltages with a separate common are obtained as follows:

1.Select the Tracking Series operating mode and set A/B Metering switch to monitor the “A” supply.

2.Set the desired voltage using the “A” VOLTAGE controls.

3.Connect a ground wire between the “A” supply negative terminal and the GND (green) terminal.

4.Turn off the power supply and the equipment to be powered during hook-up.

5.Connect the positive polarity input of the circuit to be powered to the positive (red) terminal of the “A” supply and connect the negative polarity of the circuit to the negative terminal of the “B” supply. Connect the circuit ground to the ground terminal of the “A” supply, the positive terminal of the “B” supply, or the GND (green) terminal.

35

Fig. 15. Typical Hook-Up Using Identical Positive and Negative Voltages with a Separate Common.

Image 39
Contents Instruction Manual Test Instrument Safety DC Power Supply Table of Contents Introduction Introduction Features HOOK-UP Cables Isolated OutputOverload Protection Reverse Polarity ProtectionB Supplies SpecificationsGeneral Fixed 5V SupplySupply Controls and Indicators Controls and IndicatorsControls and Indicators Constant Current Indicator Line Voltage Select Switches. Combination Fixed 5V Supply Terminals and Indicator Rear Panel ControlsRear panel controls Independent USE of a or B Supply Safety PrecautionsEquipment Precautions Operating InstructionsOperating Instructions Hook-up Operating Instructions Independent operation grounding possibilities sheet 2 Setting Current Limit Setting Current limit Series Tracking Operation Constant voltage/constant current characteristicOperating Instructions Operating Instructions Operating Instructions Parallel Tracking Operation Parallel tracking operation grounding possibilities sheet 1 Parallel tracking operation grounding possibilities sheet 2 Fixed 5 V Power Supply Operation Grounding possibilities for fixed 5 V power supply sheet 1 Grounding possibilities for fixed 5 V power supply sheet 2 Operating Instructions Applications Electronics ServicingElectronics Manufacturing Split Supply Battery ChargingElectronics Design LAB Electronics EducationPage Page Page Page Page Page Fuse Replacement MaintenanceService Information Limited One-Year Warranty Test Instrument Safety Manual DE Instrucciónes Fuente DE Poder DE CD Triple Seguridad DEL Instrumento DE Prueba Advertencia Instrucciones DE Operación Seguridad DEL Instrumento DE PruebaIntroduccion Característica DE Rastreo Unica Salida TripleFuente Fija DE DOS Fuentes DE O aSalida Aislada Protección Contra SobreflujoProtección DE Polaridad Inversa Cables Fuentes a Y B EspecificacionesGeneral Controles E Indicadores Figura 1. Controles e indicadores del panel frontal Controles E Indicadores DE LA Fuente B Indicador Constant CurrentControles DEL Panel Trasero Indicadores Y Controles DE LA Fuente DEFigura 2. Controles del panel trasero USO Independiente DE LA Fuente a O B Precauciones Para SeguridadInstrucciones DE Operación Page Page Figura 4. Operación de voltaje constante típica Operación de corriente constante típica Operación DE Rastreo EN Serie Característica de voltaje constante/corriente constantePage Page Instrucciones DE Operación Operación DE Rastreo EN Paralelo Page Page Operación DE LA Fuente Fija DE Page Page Instrucciones DE Operación Manufactura Electrónica Servicio ElectrónicoFuente Dividida Laboratorio DE Diseño ElectrónicoEducación EN Electrónica Carga DE BateriasAplicaciones Aplicaciones Aplicaciones Aplicaciones Aplicaciones Aplicaciones Mantenimiento Información de Servicio Garantía Limitada de Un Ano Seguridad DE EL Equipo DE Prueba Declaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA 2005 B&K Precision Corp