Teledyne THCD-100 Front Panel Screens, Main Screen not changing setpoint, Switch Functions

Page 12

4.3.Front Panel Screens

Here follows a complete breakdown of all THCD-100 screens:

a)Main Screen (not changing setpoint)

The main screen comes in two forms. The first shows the main channel input data (with range, fullscale and calibration applied) and the associated units string. If an input is over range (more than 15% above full scale voltage) then the data value is replaced with the over range error – “RANGE!”. This screen also shows the current mode of the setpoint, if not in AUTO (i.e. OPEN or CLOSE). If either of the relays have been tripped then they are also indicated as such on the far right of the screen.

Switch Functions:

 

Func

No function *

Go to Main Menu screen

Enable setpoint mode override, active for 2 seconds

(override mode)

Change setpoint mode to AUTO

Change setpoint value (main screen changes as below)

(override mode)

Change setpoint mode to OPEN

Change setpoint value (main screen changes as below)

(override mode)

Change setpoint mode to CLOSE

Zero

Perform input rezero (if held for 3 seconds)

*By default, the Func switch does nothing while this screen is displayed.

b)Main Screen (changing setpoint)

The other form of main screen is shown when changing the live setpoint (via an initial press of the ‘’ or ‘’ switches). All the information shown on the first main screen is shown here (albeit some items are resized or repositioned) but in addition the current setpoint value is also shown. Note that this screen is not permanent – if the ‘’ or ‘’ switches are not pressed at all for 5 seconds, then the main screen reverts to the first variety shown previously.

Switch Functions:

Func

No function*(See (a) above)

Go to Main Menu screen

Enable setpoint mode override, active for 2 seconds

(override mode)

Change setpoint mode to AUTO

Change setpoint value (hold for continuous change)

(override mode)

Change setpoint mode to OPEN

Change setpoint value (hold for continuous change

(override mode)

Change setpoint mode to CLOSE

Zero

Perform input rezero (if held for 3 seconds)

165-072011 THCD-100 Power Supply

Page 12 of 28

Image 12
Contents Teledyne Hastings Manual Print History Table of Contents Features SpecificationsGeneral Information Installation Transducer socket Pinout 15 Pin D TypePin Number Designation Interface Socket Pin out Miniature 15 Pin D Type InterfaceSerial Socket Pinout 9 Pin D Type Pin Number Designation 1Power Connection Socket Panel MountingFeatures IntroductionAnalog Inputs Analog OutputsSetpoint Mode User Re-zeroAdaptive Filtering Setpoint SourceFilter Size Common Switch Functions Edit ModeMenu Screens Editable ScreensSwitch Functions Front Panel ScreensMain Screen not changing setpoint Main Screen changing setpointSwitch Functions Main Menu ScreenSetpoint Screen Input Screen Filtering ScreenComms Menu Screen Serial ScreenFactory Defaults Confirmation Screen Relay ScreenCommand/Query Format Serial CommunicationMain Commands Output Current Readings ‘r’Output Current Readings Repeatedly ‘rp’ Setpoint Mode ‘spm’ Setpoint Value ‘spv’Setpoint Source ‘sps’ Setpoint Initial Value ‘siv’Serial Commands Setpoint Initial Mode ‘sim’Serial Baud Rate ‘bra’ Serial Communication Protocol ‘pro’Channel Setup Commands Filtering CommandsRelay Control Commands Filter Size ‘fls’Relay Trip Point ‘rlt’ Relay Hysteresis ‘rlh’Other Commands Retrieve All Settings ‘ras’User Input Re-zero ‘irz’ Date of Last Factory Calibration ‘dlc?’Service & Calibration ServiceCleaning End of LifeAppendices Appendix aSetting Default Value Additional Information Appendix BSettings in ‘ras’ command string Appendix CDescription Type and size Warranty Warranty Repair PolicyNon-Warranty Repair Policy