Teledyne THCD-100 User Re-zero, Adaptive Filtering, Setpoint Mode, Setpoint Source

Page 9

(assuming setpoint is in AUTO mode and not a slave source), then the output voltage of the setpoint would be 0.5V.

The setpoint does have some configuration parameters that can be used to alter the function of the setpoint control and they are detailed as follows:

a)Setpoint Mode

The setpoint can be configured in one of three modes – AUTO, OPEN and CLOSED.

In AUTO mode the setpoint output is dependant on the setpoint source and value settings (see below).

In OPEN mode the setpoint outputs a voltage greater than the full scale of the

device. Given that the setpoint command full scale is 5V, the output in this mode is 7V.

In CLOSED mode the setpoint outputs a voltage less than the minimum output voltage of most devices, the setpoint output voltage is –0.25V.

b)Setpoint Source

The setpoint source dictates where the setpoint value comes from, assuming the mode is set to AUTO (see above). This can be one of two possibilities – Internal or Slave.

Internal source – the setpoint uses the value set internally via the ‘& switches or via external communications.

Slave source – in this case the setpoint uses a percentage of an externally produced value. The percentage is set via the ’&‘switches and the external value comes from the secondary input channel (as mentioned 5.2 in above)

3.5.User Re-zero

Over time it is possible that the input may ‘drift’ slightly due to various conditions (temperature changes, etc.). As such it may be necessary for the channel to be re-zeroed by the user.

The user re-zero function is provided for this task. It simply takes the current reading (sampled and averaged over 3 seconds) and uses that as an additional offset for the channel in question, subtracting the value from all subsequent readings. Ensure any process value to be zeroed, is in fact truly zero before performing this function. This would mean isolating flow devices or fully pumping a pressure device.

Note that the user re-zero via serial communications also provides the facility for clearing any user re-zero value that may have already been set. This should be used before any input calibration is performed to ensure that the calculated calibration points are not distorted by the user zero offset.

3.6.Adaptive Filtering

The THCD-100 includes an adaptive averaging filter on the display and serial communication readings output to aid in ‘smoothing out’ unwanted ‘noise’ on the displayed readings.

a)Operational Band

The filter can be configured to only operate within a certain band, meaning that excursions between subsequent readings that fall outside that band are shown as real (and not filtered) readings. The band is configurable between 0.01% and 1.00% of the full scale range of the input channel. It can also be set to OFF, which means that no filtering is performed (i.e. every reading is a ‘real’ one) and can also be set to ON which means that filtering is always shown no matter how big of an excursion occurs.

165-072011 THCD-100 Power Supply

Page 9 of 28

Image 9
Contents Teledyne Hastings Manual Print History Table of Contents Features SpecificationsGeneral Information Installation Transducer socket Pinout 15 Pin D TypePin Number Designation Interface Interface Socket Pin out Miniature 15 Pin D TypeSerial Socket Pinout 9 Pin D Type Pin Number Designation 1Panel Mounting Power Connection SocketIntroduction FeaturesAnalog Inputs Analog OutputsUser Re-zero Setpoint ModeAdaptive Filtering Setpoint SourceFilter Size Edit Mode Common Switch FunctionsMenu Screens Editable ScreensFront Panel Screens Switch FunctionsMain Screen not changing setpoint Main Screen changing setpointSwitch Functions Main Menu ScreenSetpoint Screen Filtering Screen Input ScreenSerial Screen Comms Menu ScreenRelay Screen Factory Defaults Confirmation ScreenSerial Communication Command/Query FormatMain Commands Output Current Readings ‘r’Output Current Readings Repeatedly ‘rp’ Setpoint Value ‘spv’ Setpoint Mode ‘spm’Setpoint Source ‘sps’ Setpoint Initial Value ‘siv’Setpoint Initial Mode ‘sim’ Serial CommandsSerial Baud Rate ‘bra’ Serial Communication Protocol ‘pro’Filtering Commands Channel Setup CommandsFilter Size ‘fls’ Relay Control CommandsRelay Trip Point ‘rlt’ Relay Hysteresis ‘rlh’Retrieve All Settings ‘ras’ Other CommandsUser Input Re-zero ‘irz’ Date of Last Factory Calibration ‘dlc?’Service Service & CalibrationCleaning End of LifeAppendix a AppendicesAppendix B Setting Default Value Additional InformationSettings in ‘ras’ command string Appendix CDescription Type and size Warranty Warranty Repair PolicyNon-Warranty Repair Policy