Teledyne THCD-100 Interface Socket Pin out Miniature 15 Pin D Type, Pin Number Designation 1

Page 6

2.3.Interface Socket Pin out – Miniature 15 Pin 'D' Type

 

 

Pin Number

 

Designation

 

 

1

 

Analog Output

 

 

2

 

Relay 2 Common

 

 

3

 

Relay 2 Normally Open contact

 

 

4

 

Relay 2 Normally Closed contact

 

 

5

 

0V (Signal Ground)

 

 

6

 

External Setpoint Return

 

 

7

 

N/C

 

 

8

 

External Setpoint

 

 

9

 

N/C

 

 

10

 

0V (Signal Ground)

 

 

11

 

TTL Input – Disable Menu

 

 

12

 

TTL Input – Disable Zero

 

 

13

 

Relay 1 Common

 

 

14

 

Relay 1 Normally Open contact

 

 

15

 

Relay 1 Normally Closed contact

Interface

2.4.Serial

 

 

 

 

 

Socket Pinout – 9 Pin 'D' Type

 

 

 

 

 

 

 

 

 

 

 

 

Pin Number

 

Designation 1 2

 

 

1

 

RS485 Tx +

 

 

 

 

 

 

 

2

 

RS232 Rx

 

 

 

 

 

 

 

3

 

RS232 Tx

 

 

 

 

 

 

 

4

 

RS485 Tx –

 

 

 

 

 

 

 

5

 

Common

 

 

 

 

 

 

 

6

 

RS485 Rx +

 

 

 

 

 

 

 

7

 

RS485 Rx –

 

 

 

 

 

 

 

8

 

N/C

 

 

 

 

 

 

 

9

 

Common

 

 

 

 

 

 

 

 

1For RS232 only, connect only the pins designated in bold font.

2For RS485 only, connect only the pins designated in normal font. See section 4 for instructions on changing the serial protocol.

165-072011 THCD-100 Power Supply

Page 6 of 28

Image 6
Contents Teledyne Hastings Manual Print History Table of Contents Features SpecificationsGeneral Information Installation Transducer socket Pinout 15 Pin D TypePin Number Designation Serial Socket Pinout 9 Pin D Type Interface Socket Pin out Miniature 15 Pin D TypeInterface Pin Number Designation 1Power Connection Socket Panel MountingAnalog Inputs FeaturesIntroduction Analog OutputsAdaptive Filtering Setpoint ModeUser Re-zero Setpoint SourceFilter Size Menu Screens Common Switch FunctionsEdit Mode Editable ScreensMain Screen not changing setpoint Switch FunctionsFront Panel Screens Main Screen changing setpointSwitch Functions Main Menu ScreenSetpoint Screen Input Screen Filtering ScreenComms Menu Screen Serial ScreenFactory Defaults Confirmation Screen Relay ScreenCommand/Query Format Serial CommunicationMain Commands Output Current Readings ‘r’Output Current Readings Repeatedly ‘rp’ Setpoint Source ‘sps’ Setpoint Mode ‘spm’Setpoint Value ‘spv’ Setpoint Initial Value ‘siv’Serial Baud Rate ‘bra’ Serial CommandsSetpoint Initial Mode ‘sim’ Serial Communication Protocol ‘pro’Channel Setup Commands Filtering CommandsRelay Trip Point ‘rlt’ Relay Control CommandsFilter Size ‘fls’ Relay Hysteresis ‘rlh’User Input Re-zero ‘irz’ Other CommandsRetrieve All Settings ‘ras’ Date of Last Factory Calibration ‘dlc?’Cleaning Service & CalibrationService End of LifeAppendices Appendix aSetting Default Value Additional Information Appendix BSettings in ‘ras’ command string Appendix CDescription Type and size Warranty Warranty Repair PolicyNon-Warranty Repair Policy