f)Comms Menu Screen

The Comms Menu screen provides options for setting the serial communication’s baud rate, address and RS232/RS485 protocol.

Switch Functions:

Func

Go back to Main Menu screen

Go to previous menu selection

Go to next menu selection

No function

No function

Zero

Go to screen associated with menu selection

g)Serial Screen

The main configuration item on the serial screen allows you to select the baud rate of the THCD- 100 for serial communication. The allowed baud rates are 9600, 19200 & 57600. There is also an address field that is mainly used for identifying a particular unit if several are connected together on an RS485 network. All commands sent to the unit must be prefixed with the address letter (see section 5.2 for more information). Valid letters are ‘a’ to ‘h’. The other editable field is used to tell the THCD-100 whether it is on an RS485 serial network.

Switch Functions:

Func

Go back to Main Menu screen

Select previous character when in edit mode

Select next character when in edit mode

Move edit field selection up

Move edit field selection down

Zero

Go to edit mode for currently selected field

165-072011 THCD-100 Power Supply

Page 15 of 28

Image 15
Contents Teledyne Hastings Manual Print History Table of Contents Features SpecificationsGeneral Information Installation Transducer socket Pinout 15 Pin D TypePin Number Designation Pin Number Designation 1 Interface Socket Pin out Miniature 15 Pin D TypeInterface Serial Socket Pinout 9 Pin D TypePanel Mounting Power Connection SocketAnalog Outputs FeaturesIntroduction Analog InputsSetpoint Source Setpoint ModeUser Re-zero Adaptive FilteringFilter Size Editable Screens Common Switch FunctionsEdit Mode Menu ScreensMain Screen changing setpoint Switch FunctionsFront Panel Screens Main Screen not changing setpointSwitch Functions Main Menu ScreenSetpoint Screen Filtering Screen Input ScreenSerial Screen Comms Menu ScreenRelay Screen Factory Defaults Confirmation ScreenSerial Communication Command/Query FormatMain Commands Output Current Readings ‘r’Output Current Readings Repeatedly ‘rp’ Setpoint Initial Value ‘siv’ Setpoint Mode ‘spm’Setpoint Value ‘spv’ Setpoint Source ‘sps’Serial Communication Protocol ‘pro’ Serial CommandsSetpoint Initial Mode ‘sim’ Serial Baud Rate ‘bra’Filtering Commands Channel Setup CommandsRelay Hysteresis ‘rlh’ Relay Control CommandsFilter Size ‘fls’ Relay Trip Point ‘rlt’Date of Last Factory Calibration ‘dlc?’ Other CommandsRetrieve All Settings ‘ras’ User Input Re-zero ‘irz’End of Life Service & CalibrationService CleaningAppendix a AppendicesAppendix B Setting Default Value Additional InformationSettings in ‘ras’ command string Appendix CDescription Type and size Warranty Warranty Repair PolicyNon-Warranty Repair Policy