Teledyne THCD-100 Installation, Transducer socket Pinout 15 Pin D Type, Pin Number Designation

Page 5

2.Installation and Interconnections

2.1.Installation

The THCD-100 can be panel mounted, or used on a bench top and is a standard 96 x 48mm panel meter enclosure. Note that the instrument must always be situated in such a way as to enable adequate air circulation about the unit. If a transducer was supplied with the THCD-100 a suitable cable may have been included.

See Section 2.6 for a cut out detail for panel mounting.

2.2.Transducer socket Pinout – 15 Pin 'D' Type

Pin Number

Designation

5

Signal Common

6

Signal In (VDC)

7

Case Gnd

9

-15 VDC

11

+15 VDC

12

Valve Return

13

+24 VDC

14

Set Point Out (VDC)

All other pins

Not Connected

If the transducer is being powered from another source (e.g. a mains powered supply) it will only be necessary to connect to the two signal pins. Note: The THCD-100 has a fully floating differential input. For single ended use (i.e. transducers with differential output) join 0V (pin5) & Signal Return (pin12) at the transducer. Ensure the common mode input range is not exceeded.

165-072011 THCD-100 Power Supply

Page 5 of 28

Image 5
Contents Teledyne Hastings Manual Print History Table of Contents General Information FeaturesSpecifications Pin Number Designation InstallationTransducer socket Pinout 15 Pin D Type Interface Interface Socket Pin out Miniature 15 Pin D TypeSerial Socket Pinout 9 Pin D Type Pin Number Designation 1Panel Mounting Power Connection SocketIntroduction FeaturesAnalog Inputs Analog OutputsUser Re-zero Setpoint ModeAdaptive Filtering Setpoint SourceFilter Size Edit Mode Common Switch FunctionsMenu Screens Editable ScreensFront Panel Screens Switch FunctionsMain Screen not changing setpoint Main Screen changing setpointSetpoint Screen Switch FunctionsMain Menu Screen Filtering Screen Input ScreenSerial Screen Comms Menu ScreenRelay Screen Factory Defaults Confirmation ScreenSerial Communication Command/Query FormatOutput Current Readings Repeatedly ‘rp’ Main CommandsOutput Current Readings ‘r’ Setpoint Value ‘spv’ Setpoint Mode ‘spm’Setpoint Source ‘sps’ Setpoint Initial Value ‘siv’Setpoint Initial Mode ‘sim’ Serial CommandsSerial Baud Rate ‘bra’ Serial Communication Protocol ‘pro’Filtering Commands Channel Setup CommandsFilter Size ‘fls’ Relay Control CommandsRelay Trip Point ‘rlt’ Relay Hysteresis ‘rlh’Retrieve All Settings ‘ras’ Other CommandsUser Input Re-zero ‘irz’ Date of Last Factory Calibration ‘dlc?’Service Service & CalibrationCleaning End of LifeAppendix a AppendicesAppendix B Setting Default Value Additional InformationDescription Type and size Settings in ‘ras’ command stringAppendix C Non-Warranty Repair Policy WarrantyWarranty Repair Policy