Emerson TCL S1 Zero Cal Sensor zero failed Press Exit, Calibrate?, Sensor Output

Page 54

MODEL TCL

S1 Zero Cal Sensor zero failed

Press Exit

SECTION 9

CALIBRATION

If the zero current is much larger than expected, the zero calibration failure screen appears. The analyzer will not update the zero current. For troubleshooting assistance, see Section 11.5.

8. To return to the main display, press MENU then EXIT.

9.3.3 Procedure — Calibrating the sensor

1.If the sensor was just zeroed, place the sensor back in the flow cell. Confirm that excess sample is flowing down the inside tube of the overflow sampler. Also, verify that reagent is being delivered to the sample and that the air pump is working.

2.Adjust the chlorine concentration until it is near the upper end of the control range. Wait until the analyzer reading is stable before starting the calibration. When the TCL is first started up or when a new sensor is put in service, allow at least 30 minutes for the reading to stabilize.

Calibrate?

Sensor 1

Sensor 2

Output 1

Output 2

3.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

4.Choose the sensor you wish to calibrate. Sensor 2 appears only if you have a dual input 1056 analyzer.

S1 Calibration

Total Chlorine

Temperature

S1 Calibration ZeroZeroCalCal

In Process Cal

5.Choose Total Chlorine.

6.Choose In Process Cal

46

Image 54
Contents Total Chlorine Analyzer Read this page Before Proceeding Electrical Shock HazardPage Hazardous Area Installation Rev. Level DateModel TCL Total Chlorine Analyzer Table of Contents Contd List of TablesList of Figures IiiPage Section Description and Specifications Model TCL Sample Conditioning SystemModel 1056 Chlorine Analyzer Model 499A CL-02 Sensor General Sample RequirementsSample Conditioning System Electrolyte Capacity Approximately 25 mL Linearity per ISO 15839 0-10 ppm 2% 0-20 ppm 3%Ordering Information and Accessories Accessories Sample Conditioning SystemDescription Weight Ship Weight DescriptionSchematic of Sample Conditioning System and Analyzer Principles of OperationInstallation InstallationUnpacking and Inspection Panel Mounting Dimensions Wall / Surface Mount Pipe MountInstall the Sample Conditioning Enclosure Install the SensorTCL Case Dimensions Wiring Provide Power to the Sample Conditioning SystemPrepare Analyzer Conduit Openings Power Analog output wiringSensor Wiring Alarm wiringTotal Chlorine Apply Power to the Analyzer and Complete Quick StartEnglish Ppm mg/LTemp Units Menu Tree Zero the Sensor START-UPPrepare the Reagent Start Sample FlowDisplay and Operation DisplayKeypad Outputs Programming the ANALYZER-TUTORIALCalibrate RangeO2 S1 4mA 0.0C O2 S1 20mA 100.0C Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm SecurityUsing Hold HoldLanguage English Contrast Configuring the Main DisplayMain Format Format, Language, Warning, and ContrastDefault Settings Programming the AnalyzerDefault Settings Choices Measurement ChlorineTemperature related settings Default SettingsChoices Default Security CodeCONFIGURING, RANGING, and Simulating Outputs Outputs FirstDefinitions Configure ProgramOutputAssign Output Output Range Configure Simulate Output ConfigureProcedure Ranging Outputs O1 S1 20mA 10.00 ppm O2 S1 4mA O2 S1 20mARangeSimulate Configure SimulateConfiguring Alarms and Assigning Setpoints OutputDefinitions Alarms Configure/SetpointAlarm Simulate Alarms Configure/SetpointAlarm Simulate Alarm Don’t SimulateConfiguring the Measurement Alarms Configure/Setpoint SimulateSynch Timers Yes Procedure Synchronizing TimersMeasurement Sensor1SensorTemperature Configuring Temperature Related SettingsProgram Outputs Alarms Measurement Temperature UnitsUnitsC S1 Temp Comp AutoSecurity Configuring Security SettingsProgram Alarms Measurement Temperature SecurityReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerDigital Communications HartCalibrating Temperature CalibrationIntroduction Calibrate SensorSensor11 Output Temperature S1 Calibration + 25.0CProcedure S1 CalibrationCalibrating Total Chlorine Determination of Total ChlorineProcedure Zeroing the Sensor Choose Total ChlorineCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalS1 Zero Cal Sensor zero failed Press Exit SensorOutput Calibrating Analog OutputsCalibrate? Sensor MA Output Cal Meter 000 mAOutput Trim Complete Maintenance AnalyzerTotal Chlorine Sensor Cleaning the membraneReplacing the membrane and electrolyte solution GeneralSensor Parts Spare Parts Reagent Sample and reagent tubingReplacing Sample Tubing Peristaltic pump tubing Model TCL Replacing the air pump Replacing the air pump diaphragm and check valves Replacement Parts and Reagent for Sample Conditioning System DescriptionFaults Section TroubleshootingUsing the Diagnostic Feature OverviewTroubleshooting When a Fault Message is Showing Fault message Explanation SectionHardware Error Sensor Incompatible Sensor CPU ErrorSensor ADC Error Sensor RTD OpenSensor Need Factory Cal Troubleshooting When a Warning Message is ShowingExplanation Section Sensor Negative ReadingZero current is too high Troubleshooting When no Error Message is ShowingProblem See Section Zero current is unstableSensitivity is low or readings are low Readings drift Simulating Inputs AnodCath Simulating Temperature Simulating temperatureWarranty Repair NON-WARRANTY RepairSection Return of Material Warranty Return of MaterialSpecifications subject to change without notice

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.