Emerson TCL instruction manual Sensitivity is low or readings are low

Page 72

MODEL TCL

SECTION 11

 

TROUBLESHOOTING

11.5.3 Sensitivity is low or readings are low

1.Does the reagent carboy contain reagent? Is the reagent uptake tubing below the level of the reagent? Has potassium iodide been added to the acetic acid (vinegar) reagent?

2.Is there adequate flow to the overflow sampler? Excess sample should be flowing down the inside tube of the overflow sampler.

3.Does the reagent contain the correct amount of potassium iodide? See the table.

 

Expected range,

Amount of KI needed

Part number

 

ppm as Cl2

per 5 gallons of vinegar

 

 

 

 

 

 

 

 

 

0 – 5 ppm

25 grams

24164-00

 

 

 

 

 

 

 

0 – 10 ppm

50 grams

24164-01

 

 

 

 

 

 

 

0 – 20 ppm

2 x 50 grams

24164-01

 

4.Was the comparison or calibration sample tested as soon as it was taken? Chlorine solutions can be unsta- ble. Test the sample immediately after collecting it. Avoid exposing the sample to sunlight.

5.Is the membrane fouled or coated? A dirty membrane inhibits diffusion of iodine through the membrane, reduc- ing sensor current. Clean the membrane by rinsing it with a stream of water from a wash bottle. Wipe gently with a soft tissue.

6.Are the reagent and sample pumps running? If a pump is not running, check the fuse and replace it if neces- sary. See Table 10-2 for part numbers. If the fuse is okay, replace the pump.

7.Are all tube fittings tight? Pay particular attention to the luer fittings that connect the tubing to the pumps.

8.Does the pump tubing element need replacing? Remove the tubing from the pump and inspect it. If the tubing appears permanently pinched or deformed, replace the tubing. Refer to Section 10.3.3 for instructions on how to remove and replace the tubing elements. The expected life of a tubing element is about one year.

9.Is the sample flow to the sensor about 11 mL/min? If the sample flow is too low, the total chlorine reading will be low. If the flow is too high, the ratio between the sample flow and reagent flow will be too high, and there might be insufficient reagent to properly react with the total chlorine in the sample. To check sample flow…

a.Turn off the reagent and sample pumps.

b.Disconnect the luer fitting on the discharge of the sample pump. See A in Figure 11-1.

c.Hold a small beaker under the discharge port.

d.Start the sample pump and collect sample for two minutes.

e.Measure the volume of sample collected in the beaker. After two minutes, the volume should be about 22 mL.

10.Is the reagent flow about 0.2 mL/min? If the reagent flow is too low, there might be insufficient acetic acid to lower the sample pH and insufficient potassium iodide to react with total chlorine in the sample. To check reagent flow…

a.Turn off the reagent and sample pumps.

b.Disconnect the reagent tubing at the injection tee. See B in Figure 11-1.

c.Place the end of the tubing in a 5 mL graduated cylinder.

d.Start the reagent pump and collect reagent for ten minutes.

e.Note the volume of reagent collected in the graduated cylinder. After ten minutes the volume should be about 2 mL.

64

Image 72
Contents Total Chlorine Analyzer Read this page Before Proceeding Electrical Shock HazardPage Hazardous Area Installation Rev. Level DateModel TCL Total Chlorine Analyzer Table of Contents Contd List of TablesList of Figures IiiPage Section Description and Specifications Model TCL Sample Conditioning SystemModel 1056 Chlorine Analyzer Model 499A CL-02 Sensor General Sample RequirementsSample Conditioning System Electrolyte Capacity Approximately 25 mL Linearity per ISO 15839 0-10 ppm 2% 0-20 ppm 3%Ordering Information and Accessories Accessories Sample Conditioning SystemDescription Weight Ship Weight DescriptionSchematic of Sample Conditioning System and Analyzer Principles of OperationInstallation InstallationUnpacking and Inspection Panel Mounting Dimensions Wall / Surface Mount Pipe MountInstall the Sample Conditioning Enclosure Install the SensorTCL Case Dimensions Wiring Provide Power to the Sample Conditioning SystemPrepare Analyzer Conduit Openings Power Analog output wiringSensor Wiring Alarm wiringApply Power to the Analyzer and Complete Quick Start EnglishTotal Chlorine Ppm mg/LTemp Units Menu Tree START-UP Prepare the ReagentZero the Sensor Start Sample FlowDisplay and Operation DisplayKeypad Programming the ANALYZER-TUTORIAL CalibrateOutputs RangeOutput Range O1 S1 4mA 0.000 ppm O1 S1 20mA 08.50 ppmO2 S1 4mA 0.0C O2 S1 20mA 100.0C SecurityUsing Hold HoldConfiguring the Main Display Main FormatLanguage English Contrast Format, Language, Warning, and ContrastDefault Settings Programming the AnalyzerDefault Settings Choices Measurement ChlorineDefault Settings Choices DefaultTemperature related settings Security CodeCONFIGURING, RANGING, and Simulating Outputs Outputs FirstDefinitions Configure ProgramOutputAssign Configure Simulate Output Configure Procedure Ranging OutputsOutput Output Range O1 S1 20mA 10.00 ppm O2 S1 4mA O2 S1 20mAConfigure Simulate Configuring Alarms and Assigning SetpointsRangeSimulate OutputDefinitions Alarms Configure/SetpointAlarm Alarms Configure/Setpoint Alarm Simulate AlarmSimulate Don’t SimulateAlarms Configure/Setpoint Simulate Synch Timers YesConfiguring the Measurement Procedure Synchronizing TimersMeasurement Sensor1SensorConfiguring Temperature Related Settings Program Outputs Alarms MeasurementTemperature Temperature UnitsUnitsC S1 Temp Comp AutoConfiguring Security Settings Program Alarms Measurement TemperatureSecurity SecurityResetting the Analyzer Program Temperature Security DiagnosticsReset Analyzer Reset AnalyzerDigital Communications HartCalibrating Temperature CalibrationIntroduction Temperature S1 Calibration + 25.0C ProcedureCalibrate SensorSensor11 Output S1 CalibrationCalibrating Total Chlorine Determination of Total ChlorineProcedure Zeroing the Sensor Choose Total ChlorineTemperature S1 Calibration ZeroZeroCalCal Process Cal S1 Zero Cal Sensor zero failed Press ExitCalibrate? SensorCalibrating Analog Outputs Calibrate? SensorOutput MA Output Cal Meter 000 mAOutput Trim Complete Maintenance AnalyzerCleaning the membrane Replacing the membrane and electrolyte solutionTotal Chlorine Sensor GeneralSensor Parts Spare Parts Reagent Sample and reagent tubingReplacing Sample Tubing Peristaltic pump tubing Model TCL Replacing the air pump Replacing the air pump diaphragm and check valves Replacement Parts and Reagent for Sample Conditioning System DescriptionSection Troubleshooting Using the Diagnostic FeatureFaults OverviewTroubleshooting When a Fault Message is Showing Fault message Explanation SectionHardware Error Sensor CPU Error Sensor ADC ErrorSensor Incompatible Sensor RTD OpenTroubleshooting When a Warning Message is Showing Explanation SectionSensor Need Factory Cal Sensor Negative ReadingTroubleshooting When no Error Message is Showing Problem See SectionZero current is too high Zero current is unstableSensitivity is low or readings are low Readings drift Simulating Inputs AnodCath Simulating Temperature Simulating temperatureWarranty Repair NON-WARRANTY RepairSection Return of Material Warranty Return of MaterialSpecifications subject to change without notice

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.