Press Exit

SECTION 9

CALIBRATION

7.Follow the screen prompts: Once the reading is stable, press ENTER. Take the sample and press ENTER. At this point, the analyzer will store the present sensor current and temperature and use those values in the calibration.

Determine the total chlorine concentration in the sample and enter the value in the screen shown at left. See Section 9.3.1 for sampling and testing precautions.

8.If the calibration is successful, the live reading will change to the value entered in step 7 and the display will return to the screen in step 6.

If the sensitivity is too far outside the range of expected values, the calibration error screen shown at left will appear. The analyzer will not update the calibration. For troubleshooting assistance, see Section 11.5.

9. To return to the main display, press MENU then EXIT.

9.4 CALIBRATING ANALOG OUTPUTS

9.4.1 Purpose

Although the analyzer analog outputs are calibrated at the factory, they can be trimmed in the field to match the reading from a standard milliammeter. Both the low (0 or 4 mA) and high (20 mA) outputs can be trimmed.

9.4.2 Procedure

1.Connect a calibrated milliammeter across the output you wish to calibrate. If a load is already connected to the output, disconnect the load. Do not put the milliameter in parallel with the load.

Calibrate?

Sensor 1

Sensor 2

Output 1

Output 2

2.Press MENU. The main menu screen appears. The cursor will be on Calibrate. Press ENTER.

3.Choose the output you wish to calibrate.

4 mA Output 1

Cal Meter:

0 4.000 mA

20 mA Output 1

Cal Meter:

2 0.000 mA

4.The analyzer will simulate the low output current. Change the value in the display to match the reading from the milliammeter.

5.The analyzer will simulate the 20 mA output current. Change the value in the display to match the reading from the milliammeter.

47

Image 55
Contents Total Chlorine Analyzer Electrical Shock Hazard Read this page Before ProceedingPage Rev. Level Date Hazardous Area InstallationModel TCL Total Chlorine Analyzer List of Tables Table of Contents ContdIii List of FiguresPage Model TCL Sample Conditioning System Section Description and SpecificationsModel 1056 Chlorine Analyzer Model 499A CL-02 Sensor Sample Requirements GeneralSample Conditioning System Linearity per ISO 15839 0-10 ppm 2% 0-20 ppm 3% Electrolyte Capacity Approximately 25 mLAccessories Sample Conditioning System Ordering Information and AccessoriesDescription Description Weight Ship WeightPrinciples of Operation Schematic of Sample Conditioning System and AnalyzerInstallation InstallationUnpacking and Inspection Panel Mounting Dimensions Pipe Mount Wall / Surface MountInstall the Sensor Install the Sample Conditioning EnclosureTCL Case Dimensions Provide Power to the Sample Conditioning System WiringPrepare Analyzer Conduit Openings Analog output wiring PowerAlarm wiring Sensor WiringPpm mg/L Apply Power to the Analyzer and Complete Quick StartEnglish Total ChlorineTemp Units Menu Tree Start Sample Flow START-UPPrepare the Reagent Zero the SensorDisplay Display and OperationKeypad Range Programming the ANALYZER-TUTORIALCalibrate OutputsSecurity Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm O2 S1 4mA 0.0C O2 S1 20mA 100.0CHold Using HoldFormat, Language, Warning, and Contrast Configuring the Main DisplayMain Format Language English ContrastProgramming the Analyzer Default SettingsMeasurement Chlorine Default Settings ChoicesSecurity Code Default SettingsChoices Default Temperature related settingsOutputs First CONFIGURING, RANGING, and Simulating OutputsDefinitions ProgramOutput ConfigureAssign O1 S1 20mA 10.00 ppm O2 S1 4mA O2 S1 20mA Configure Simulate Output ConfigureProcedure Ranging Outputs Output Output RangeOutput Configure SimulateConfiguring Alarms and Assigning Setpoints RangeSimulateDefinitions Configure/Setpoint AlarmsAlarm Don’t Simulate Alarms Configure/SetpointAlarm Simulate Alarm SimulateProcedure Synchronizing Timers Alarms Configure/Setpoint SimulateSynch Timers Yes Configuring the MeasurementSensor1Sensor MeasurementTemperature UnitsUnitsC S1 Temp Comp Auto Configuring Temperature Related SettingsProgram Outputs Alarms Measurement TemperatureSecurity Configuring Security SettingsProgram Alarms Measurement Temperature SecurityReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerHart Digital CommunicationsCalibration Calibrating TemperatureIntroduction S1 Calibration Temperature S1 Calibration + 25.0CProcedure Calibrate SensorSensor11 OutputDetermination of Total Chlorine Calibrating Total ChlorineChoose Total Chlorine Procedure Zeroing the SensorSensor Temperature S1 Calibration ZeroZeroCalCal Process CalS1 Zero Cal Sensor zero failed Press Exit Calibrate?MA Output Cal Meter 000 mA Calibrating Analog OutputsCalibrate? Sensor OutputOutput Trim Complete Analyzer MaintenanceGeneral Cleaning the membraneReplacing the membrane and electrolyte solution Total Chlorine SensorSensor Parts Spare Parts Sample and reagent tubing ReagentReplacing Sample Tubing Peristaltic pump tubing Model TCL Replacing the air pump Replacing the air pump diaphragm and check valves Description Replacement Parts and Reagent for Sample Conditioning SystemOverview Section TroubleshootingUsing the Diagnostic Feature FaultsFault message Explanation Section Troubleshooting When a Fault Message is ShowingHardware Error Sensor RTD Open Sensor CPU ErrorSensor ADC Error Sensor IncompatibleSensor Negative Reading Troubleshooting When a Warning Message is ShowingExplanation Section Sensor Need Factory CalZero current is unstable Troubleshooting When no Error Message is ShowingProblem See Section Zero current is too highSensitivity is low or readings are low Readings drift Anod Simulating InputsCath Simulating temperature Simulating TemperatureNON-WARRANTY Repair Warranty RepairSection Return of Material Return of Material WarrantySpecifications subject to change without notice

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.