Emerson TCL instruction manual Power, Analog output wiring

Page 22

MODEL TCL

SECTION 4

 

WIRING

4.3 MAKE POWER, ALARM, OUTPUT, AND SENSOR CONNECTIONS IN THE ANALYZER

WARNING

RISK OF ELECTRICAL SHOCK

Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local code.

4.3.1 Power

Wire AC mains power to the power supply board, which is mounted vertically on the left hand side of the analyzer enclosure. The power connector is at the top of the board. Unplug the connector from the board and wire the power cable to it. Lead connections are marked on the connector. (L is live or hot; N is neutral, the ground connection has the standard symbol.)

AC power wiring should be 14 gauge or greater. Run the power wiring through the conduit opening nearest the power terminal. Provide a switch or breaker to disconnect the analyzer from the main power supply. Install the switch or breaker near the analyzer and label it as the disconnecting device for the analyzer.

CAUTION

If your 1056 analyzer does not have alarm relays (options -270, -271, or -272) you must set the black and red AC power switch located below the power terminal to the correct AC voltage. The analyzer is shipped with the switch in the 230 VAC position. For operation at 110-120 VAC, slide the switch upward so that 115 VAC is showing.

If your 1056 analyzer has alarm relays (options -273, -274, or -275) there is no switch setting to make. The analyzer automatically detects the AC voltage.

4.3.2 Analog output wiring

Two analog current outputs are located on the main circuit board, which is attached to the inside of the enclosure door. Figure 4-3 shows the location of the terminals. The connectors can be detached for wiring. TB-1 is output 1. TB-2 is output 2. Polarity is marked on the circuit board.

For best EMI/RFI protection, use shielded output signal cable enclosed in earth-grounded metal conduit.

Keep output signal wiring separate from power wiring. Do not run signal and power or relay wiring in the same conduit or close together in a cable tray.

FIGURE 4-2. Analog output connections. The analog outputs are on the main board near the hinged end of the enclosure door.

14

Image 22
Contents Total Chlorine Analyzer Read this page Before Proceeding Electrical Shock HazardPage Hazardous Area Installation Rev. Level DateModel TCL Total Chlorine Analyzer Table of Contents Contd List of TablesList of Figures IiiPage Model TCL Sample Conditioning System Section Description and SpecificationsModel 1056 Chlorine Analyzer Model 499A CL-02 Sensor Sample Requirements GeneralSample Conditioning System Electrolyte Capacity Approximately 25 mL Linearity per ISO 15839 0-10 ppm 2% 0-20 ppm 3%Ordering Information and Accessories Accessories Sample Conditioning SystemDescription Weight Ship Weight DescriptionSchematic of Sample Conditioning System and Analyzer Principles of OperationInstallation InstallationUnpacking and Inspection Panel Mounting Dimensions Wall / Surface Mount Pipe MountInstall the Sample Conditioning Enclosure Install the SensorTCL Case Dimensions Provide Power to the Sample Conditioning System WiringPrepare Analyzer Conduit Openings Power Analog output wiringSensor Wiring Alarm wiringTotal Chlorine Apply Power to the Analyzer and Complete Quick StartEnglish Ppm mg/LTemp Units Menu Tree Zero the Sensor START-UPPrepare the Reagent Start Sample FlowDisplay and Operation DisplayKeypad Outputs Programming the ANALYZER-TUTORIALCalibrate RangeO2 S1 4mA 0.0C O2 S1 20mA 100.0C Output Range O1 S1 4mA 0.000 ppmO1 S1 20mA 08.50 ppm SecurityUsing Hold HoldLanguage English Contrast Configuring the Main DisplayMain Format Format, Language, Warning, and ContrastDefault Settings Programming the AnalyzerDefault Settings Choices Measurement ChlorineTemperature related settings Default SettingsChoices Default Security CodeOutputs First CONFIGURING, RANGING, and Simulating OutputsDefinitions ProgramOutput ConfigureAssign Output Output Range Configure Simulate Output ConfigureProcedure Ranging Outputs O1 S1 20mA 10.00 ppm O2 S1 4mA O2 S1 20mARangeSimulate Configure SimulateConfiguring Alarms and Assigning Setpoints OutputDefinitions Configure/Setpoint AlarmsAlarm Simulate Alarms Configure/SetpointAlarm Simulate Alarm Don’t SimulateConfiguring the Measurement Alarms Configure/Setpoint SimulateSynch Timers Yes Procedure Synchronizing TimersMeasurement Sensor1SensorTemperature Configuring Temperature Related SettingsProgram Outputs Alarms Measurement Temperature UnitsUnitsC S1 Temp Comp AutoSecurity Configuring Security SettingsProgram Alarms Measurement Temperature SecurityReset Analyzer Resetting the AnalyzerProgram Temperature Security Diagnostics Reset AnalyzerDigital Communications HartCalibration Calibrating TemperatureIntroduction Calibrate SensorSensor11 Output Temperature S1 Calibration + 25.0CProcedure S1 CalibrationCalibrating Total Chlorine Determination of Total ChlorineProcedure Zeroing the Sensor Choose Total ChlorineCalibrate? Temperature S1 Calibration ZeroZeroCalCal Process CalS1 Zero Cal Sensor zero failed Press Exit SensorOutput Calibrating Analog OutputsCalibrate? Sensor MA Output Cal Meter 000 mAOutput Trim Complete Maintenance AnalyzerTotal Chlorine Sensor Cleaning the membraneReplacing the membrane and electrolyte solution GeneralSensor Parts Spare Parts Reagent Sample and reagent tubingReplacing Sample Tubing Peristaltic pump tubing Model TCL Replacing the air pump Replacing the air pump diaphragm and check valves Replacement Parts and Reagent for Sample Conditioning System DescriptionFaults Section TroubleshootingUsing the Diagnostic Feature OverviewFault message Explanation Section Troubleshooting When a Fault Message is ShowingHardware Error Sensor Incompatible Sensor CPU ErrorSensor ADC Error Sensor RTD OpenSensor Need Factory Cal Troubleshooting When a Warning Message is ShowingExplanation Section Sensor Negative ReadingZero current is too high Troubleshooting When no Error Message is ShowingProblem See Section Zero current is unstableSensitivity is low or readings are low Readings drift Anod Simulating InputsCath Simulating Temperature Simulating temperatureNON-WARRANTY Repair Warranty RepairSection Return of Material Warranty Return of MaterialSpecifications subject to change without notice

TCL specifications

Emerson TCL, a leader in technology and automation solutions, offers a diverse range of products and systems designed to optimize industrial processes. The company has established itself as a key player in the industry by providing innovative technologies that enhance operational efficiency, reliability, and sustainability.

One of the main features of Emerson TCL is its advanced process automation capabilities. The company's integrated solutions enable manufacturers to monitor, control, and optimize their operations in real-time. With a focus on automation, Emerson TCL helps organizations reduce operational costs, improve quality, and achieve greater throughput.

Another notable aspect of Emerson TCL is its commitment to digital transformation. The company’s technologies leverage the Internet of Things (IoT) to connect devices, systems, and people, allowing for seamless data exchange and insights. This connectivity enables operators to make informed decisions, identify trends, and implement improvements more quickly and effectively.

Emerson TCL also emphasizes the importance of safety and regulation compliance. Its solutions are designed to meet stringent industry standards, ensuring that organizations can operate within the legal framework while maintaining high safety levels. The company provides tools and software that support risk management and compliance monitoring, helping organizations avoid costly penalties and enhance their safety culture.

Sustainability is another key characteristic of Emerson TCL's offerings. The company is dedicated to helping industries reduce their environmental footprint through energy-efficient technologies and processes. By improving energy management and resource utilization, Emerson TCL supports organizations in their quest to achieve sustainability goals.

Additionally, Emerson TCL boasts a comprehensive portfolio of products, including process control systems, instrumentation, and software solutions. With a focus on modularity and scalability, the company provides customized systems that can adapt to various industry needs, making it easier for companies to implement solutions that grow with them.

In summary, Emerson TCL stands out for its advanced process automation, commitment to digital transformation, emphasis on safety, and sustainability initiatives. Through its innovative technologies and comprehensive product offerings, Emerson TCL continues to empower industries to enhance their operations and achieve their strategic goals.