Teledyne 05386D manual Calibration or Cal Check Procedure

Page 20

Model 200EU-NOy Option Manual Addendum

NOTE

The M200EU with the NOy Option does NOT have equivalency approval, and may not be used for EPA monitoring.

5.1 Calibration or Cal Check Procedure

The calibration of the instrument can be checked or adjusted using gas introduced through the calibration port “CAL IN” on the Bypass Pump Module. The calibration gas is routed to a TEE near the sample inlet port on the remote converter assembly.

Since the zero gas concentration is defined as 0 ppb, it is not necessary to enter the expected zero value. Table 3 details the zero calibration procedure.

Table 3 Manual Zero Calibration Procedure - Zero Gas Through Sample Port

Step

Action

Comment

Number

 

 

 

 

 

1.

Press CAL

The M200EU enters the calibrate mode.

 

 

 

2.

Wait For Stability

Wait for reading to stabilize at zero value.

 

Typ STB < 0.1 PPB

 

 

 

 

3.

Press ZERO

If you change your mind after pressing ZERO, you can

 

 

still press EXIT here without zeroing the instrument.

4.

Press ENTR

Pressing ENTR actually changes the calculation

 

 

equations.

5.

Press EXIT

M200EU returns to sampling. Immediately after

 

 

calibration, data is not added to the DAS averages.

 

 

 

16

05386 Rev. D

Image 20
Contents M200EU-NOy Converter Page Page Page Safety Messages 05386 Rev. D Table of Contents List of Figures List of TablesIntroduction This page intentionally left blank 05386 Rev. D Electrical and Pneumatic Connections Getting StartedUnpacking Rear Panel Pneumatic Connections Initial Operation Final Test and Calibration ValuesThis page intentionally left blank 05386 Rev. D SPECIFICATIONS, Warranty SpecificationsWarranty Teledyne Advanced Pollution INSTRUMENTATION, INCM200EU-NOYCONVERTER Principle of OperationThis page intentionally left blank 05386 Rev. D Calibration and ZERO/SPAN Checks Calibration or Cal Check Procedure Span Calibration Procedure This page intentionally left blank 05386 Rev. D Maintenance Maintenance ScheduleReplacing the Sample Particulate Filters Replacing the Particulate Filter Checking Analyzer Flow Rate Replacing the ConverterNOy Converter Assembly Inspecting Pneumatic Lines Pneumatics diagram standard and Z/S option Pneumatic diagram OPT 86 option Sample Driers Checking for Leaks This page intentionally left blank 05386 Rev. D TROUBLESHOOTING, Adjustments Operation Verification Diagnostic Techniques Pneumatic system is diagramed 05386 Rev. D Setting the Converter Temperature Bypass Pump Diagnostic ProceduresLOC Lock AH Alarm Electrical Diagram This page intentionally left blank 05386 Rev. D M200EU-NOYSPARE Parts List Description