7.1.2 Leak Check
Refer to Section 5.6 for the leak check procedure.
7.1.3 Bypass Pump Diagnostic Procedures
The bypass pump is capable of maintaining a total bypass flowrate of ~2.2 lpm. If flow checks indicate a fall off in bypass flow, the pump may need servicing. Check the pump, inlet fittings, and analyzer for leaks first. If other causes have been eliminated, rebuild the pump.
7.1.4 Electrical Fault Isolation
115 VAC is supplied to:
the bypass pump
the +5 volt and +24 volt switcher supplies
the input power to the converter temperature controller
Use Figure 6 to check the 115 VAC supply voltage to each of these components.
The converter temperature is controlled by the FUJI controller located on the Pump Module front panel. The controller cycles power to the external converter, thereby maintaining the temperature at 315 OC, using the solid state relay.
7.2 Setting the Converter Temperature
A front
1.Select SV with the PV/SV button,
2.Press the
3.(the digit will flash),
4.Press the
5.Repeat for the other two digits,
6.Press the ENTER button.
7.Select PV with the PV/SV button to observe the actual temperature.
7.2.1Temperature Controller Setup
The FUJI temperature controller has been programmed at the factory and should not need to be altered under normal usage.