Model 200EU-NOy Option Manual Addendum

Indications that would require re-programming are:

Failure / replacement of controller.

Replacement of thermocouple element.

Controller’s loss of PID parameters leading to instability of the temp setpoint.

In the event that the control parameters are changed or in the event that a new controller is installed, it must be reprogrammed to suit the thermal characteristics of the instrument. The controller includes an Autotuning feature, which can be used to properly set the PID factors.

To start the Autotune procedure:

1.Press Sel until A7 (AT) is displayed

2.Press Data until value = 1 (enable Autotune)

While Autotuning, the decimal point LED will blink.

NOTE: During the procedure, it is normal for the temp to overshoot up to 100°C.

When the decimal point LED turns off, Autotune is complete. No further adjustment should be necessary.

Table 7 Temperature Controller Programming Guide

PRESS

DISPLAY

ACTION

SEL

P PROP BAND

UP/DOWN SET TO 5.0 (ENTER)

DATA

 

 

SEL

i INTEGRAL

SET TO 50

DATA

 

 

SEL

d DERIVATIVE TIME

SET TO 15

DATA

 

 

SEL

AL LOW ALARM SETP

SET TO 300

DATA

 

 

SEL

AH HIGH ALARM SETP

SET TO 330

DATA

 

 

SEL

7C CYCLE TIME

SET TO 2 (FOR FAST SYSTEM)

DATA

HYS HYSTERESIS

 

SEL

SET TO 0.3

SEL

LOC LOCK

0 = OPEN 1 = LOCKED

 

2 = SV ONLY OPEN

 

 

NOTE:

DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC.

05386 Rev. D

33

Image 37
Contents M200EU-NOy Converter Page Page Page Safety Messages 05386 Rev. D Table of Contents List of Tables List of FiguresIntroduction This page intentionally left blank 05386 Rev. D Unpacking Getting StartedElectrical and Pneumatic Connections Rear Panel Pneumatic Connections Final Test and Calibration Values Initial OperationThis page intentionally left blank 05386 Rev. D Specifications SPECIFICATIONS, WarrantyTeledyne Advanced Pollution INSTRUMENTATION, INC WarrantyPrinciple of Operation M200EU-NOYCONVERTERThis page intentionally left blank 05386 Rev. D Calibration and ZERO/SPAN Checks Calibration or Cal Check Procedure Span Calibration Procedure This page intentionally left blank 05386 Rev. D Maintenance Schedule MaintenanceReplacing the Sample Particulate Filters Replacing the Particulate Filter Replacing the Converter Checking Analyzer Flow RateNOy Converter Assembly Inspecting Pneumatic Lines Pneumatics diagram standard and Z/S option Pneumatic diagram OPT 86 option Sample Driers Checking for Leaks This page intentionally left blank 05386 Rev. D TROUBLESHOOTING, Adjustments Operation Verification Diagnostic Techniques Pneumatic system is diagramed 05386 Rev. D Bypass Pump Diagnostic Procedures Setting the Converter TemperatureLOC Lock AH Alarm Electrical Diagram This page intentionally left blank 05386 Rev. D Description M200EU-NOYSPARE Parts List