Model 200EU-NOy Option Manual Addendum

6.2 Replacing the Sample Particulate Filters

The particulate filter should be inspected often for signs of plugging or contamination. It is also common for dirt particles to cause instrument drift, and affect accuracy.

To check and change the filter:

1.Fold down the Bypass Pump Module front panel.

2.Locate both filters on the left and right side of the Pump Module front panel. See Figure 2 for an exploded view of the filter assembly.

3.Visually inspect the filter through the glass window.

4.If the filter appears dirty, unscrew the hold-down ring, remove the Teflon o-ring and then the filter.

5.Replace the filter, being careful that the element is fully seated in the bottom of the holder. Replace the Teflon o-ring, then screw on the hold-down ring and hand tighten.

NOTE:

Handle the wetted surfaces as little as possible and clean with cloth prior to assembly

20

05386 Rev. D

Image 24
Contents M200EU-NOy Converter Page Page Page Safety Messages 05386 Rev. D Table of Contents List of Figures List of TablesIntroduction This page intentionally left blank 05386 Rev. D Getting Started UnpackingElectrical and Pneumatic Connections Rear Panel Pneumatic Connections Initial Operation Final Test and Calibration ValuesThis page intentionally left blank 05386 Rev. D SPECIFICATIONS, Warranty SpecificationsWarranty Teledyne Advanced Pollution INSTRUMENTATION, INCM200EU-NOYCONVERTER Principle of OperationThis page intentionally left blank 05386 Rev. D Calibration and ZERO/SPAN Checks Calibration or Cal Check Procedure Span Calibration Procedure This page intentionally left blank 05386 Rev. D Maintenance Maintenance ScheduleReplacing the Sample Particulate Filters Replacing the Particulate Filter Checking Analyzer Flow Rate Replacing the ConverterNOy Converter Assembly Inspecting Pneumatic Lines Pneumatics diagram standard and Z/S option Pneumatic diagram OPT 86 option Sample Driers Checking for Leaks This page intentionally left blank 05386 Rev. D TROUBLESHOOTING, Adjustments Operation Verification Diagnostic Techniques Pneumatic system is diagramed 05386 Rev. D Setting the Converter Temperature Bypass Pump Diagnostic ProceduresLOC Lock AH Alarm Electrical Diagram This page intentionally left blank 05386 Rev. D M200EU-NOYSPARE Parts List Description