Pipettor Programming
7.Attach a pipette tip to the tip cone.
The pipettor is now ready to begin the dilution. The display indicates that the pipettor is ready to aspirate the selected diluent volume.
8.Place the tip in the diluent and press the
Start Button. The unit will aspirate the diluent volume selected.
The following display indicates that the unit is ready to aspirate the selected sample volume.
9. Place the tip in the sample solution and press the Start Button to aspirate 0.50mL.
The next display indicates that the unit is ready to dispense the total volume with blowout.
10.Position the pipette tip to dispense and press the Start Button. The unit will dispense the full amount from the tip.
The display now indicates that the unit is ready to repeat the dilution sequence, beginning with the diluent volume aspiration.
Page 10 of 15 - June 2001
Pipettor Programming
Speed Selection
Speed changes are made using these four keys:
Speed "1" is slowest and Speed "5" is fastest.
1.Press to display the aspiration speed.
2.Press or to select a new aspiration speed.
3.Press when the desired aspiration speed is displayed. Dispensing speed is then displayed.
4.Press or to select a new dispensing speed.
5.Press when the desired dispensing speed is displayed. The display then returns to the diluent aspiration readout.
may not be used in the middle of any aspiration or dispensing cycle.
19 | 20 |