Specifications
Accuracy and Precision |
|
|
|
|
Accuracy and Precision |
|
|
|
|
Test Mode: d1 1:5 (5mL) dilution |
|
|
|
|
Test Action | Precision | Accuracy | ||
Aspirate 4.00mL | + | 0.6% | + | 0.8% |
- | - | |||
Aspirate 1.00mL | + | 1.6% | + | 2.1% |
- | - | |||
Dispense 5.00mL | + | 0.8% | + | 1.1% |
- | - | |||
|
|
|
|
|
Tolerances |
|
|
|
|
Tolerances |
|
|
|
|
Test Action | Range |
|
| |
Aspirate 4.00mL | 3.968 - 4.032 mL |
|
| |
Aspirate 1.00mL | 0.979 - 1.021 mL |
|
| |
Dispense 5.00mL | 4.945 - 5.055 mL |
|
| |
|
|
|
|
|
Materials |
|
|
|
|
Materials |
|
|
|
|
Part | Material |
|
| |
Tip Cone | Polypropylene |
|
| |
Ejector Collar | Polypropylene |
|
| |
Piston | Polypropylene |
|
| |
Control Handle | ABS |
|
| |
Tip Ejector Lever | POM |
|
| |
|
|
|
|
|
Page 14 of 15 - June 2001
Pipettor Tips
Genuine 3M Pipettor Tips are strongly recommended for use with the 3M Electronic Pipettor - 5mL. Other tips may adversely affect the accuracy of the instrument and may damage the tip cone and ejector mechanism.
Item No. Description
6489 5mL wide bore tip, 500/pkg, racked, sterile
6488 5mL wide bore tip, 500/pkg, racked, nonsterile
6487 5mL wide bore tip, 250/pkg, bulk, nonsterile
27 | 28 |