3M 1210 Instructions For Use, Connecting the AC Adapter, Charging the Pipettor Battery

Page 3

Instructions For Use

Only use and store the pipettor in accordance with the following operating conditions:

Condition

Range or Limit

Performance Testing Temperature

+19 to + 25°C

Mechanical Operating Temperature

+15 to + 35°C

Storage Temperature

-20 to + 40°C

Maximum Humidity

95%

Page 3 of 15 - June 2001

Instructions For Use

1.Connect the AC adapter into the connector jack at the rear of the charging stand's base.

2.Plug the AC adapter into an AC electrical outlet.

3.The green light at the top of the charging stand indicates that the stand is in operation. The light will always be on when plugged in and operating properly.

The following recommendations should always be followed for best results:

1.During aspiration, hold the pipettor vertically. For greatest accuracy, the pipettor should never be tilted at an angle greater than 15° during aspiration.

2.During aspiration, consistently place the tip no deeper than 5mm into the liquid.

3.During dispensing, position the tip at least 5mm above the surface to prevent bubbles from forming.

Connecting the AC Adapter

Be sure that the voltage rating of the AC adapter correctly corresponds to the voltage from the electrical outlet.

Do not use electrical outlets that are timed or switched.

Use only the AC adapter provided with the unit or an approved replacement from 3M.

1.

2.

 

Charging the Pipettor Battery

1.Switch the pipettor “ON” (see page 2).

2.Mount the pipettor into the charging stand. The top of the pipettor will fit snugly into the head of the charging stand. The magnetic mount will lock the pipettor into the proper charging position.

3.If the battery is new or has been allowed to fully discharge, keep the pipettor in the stand for 10 hours to fully charge the battery before continuing use.

The unit will charge when “ON” or “OFF”. Leave the pipettor “ON” and in the charging stand when not in use to retain user-selected pipettor settings.

Switching the unit “OFF”clears user-set programming and resets the pipettor to default settings.

5

6

Image 3
Contents Description Precautions Indications3M Electronic Pipettor 5mL includes Condition Range or Limit Connecting the AC AdapterCharging the Pipettor Battery Instructions For UseProgram Menu Power on Display and StartupMode Selection and Mode Recall Pipettor ProgrammingDilution 5mL Total Volume Dilution 1mL Total Volume Repeat Dispensing 1-5 repetitions of 1mL 1mL Pipetting Open Dilution Programming 5mL PipettingSpeed Selection Frozen Display TroubleshootingEr 1 Display BubblesFilter Replacement Care and MaintenanceCare and Maintenance Cleaning the Tip ConeBattery Replacement Calibration Check Mode d1, 15, 5mL dilutionCalibration Check Chart Tolerances SpecificationsPipettor Tips Accuracy and PrecisionLimited Warranty Out of Warranty ServiceLimitation of Remedies