3M 1210 instruction manual Dilution 5mL Total Volume

Page 5

Pipettor Programming

1:5 Dilution (5mL Total Volume)

1.Press repeatedly until the display reads:

2.Press to enter selection.

3.Attach a pipette tip to the tip cone.

The pipettor is now ready to begin the 1:5 dilution (5mL total volume). The display indicates that the pipettor is ready to aspirate 4.00mL.

4.Place the tip in the diluent and press the

Start Button. The unit will aspirate 4.00mL.

The following display indicates that the unit is ready to aspirate 1.00mL.

5. Place the tip in the sample solution and press the Start Button to aspirate 1.00mL.

The next display indicates that the unit is ready to dispense the total volume (5mL) with blowout.

6.Position the pipette tip to dispense

and press the Start Button. The unit will

dispense the full amount from the tip.

The display now indicates that the unit is ready to repeat the dilution sequence, beginning with the 4.00mL aspiration.

Page 5 of 15 - June 2001

Pipettor Programming

1:10 Dilution (5mL Total Volume)

1.Press repeatedly until the display reads:

2.Press to enter selection.

3.Attach a pipette tip to the tip cone.

The pipettor is now ready to begin the 1:10 dilution (5mL total volume). The display indicates that the pipettor is ready to aspirate 4.50mL.

4.Place the tip in the diluent and press the

Start Button. The unit will aspirate 4.50mL.

The following display indicates that the unit is ready to aspirate 0.50mL.

5. Place the tip in the sample solution and press the Start Button to aspirate 0.50mL.

The next display indicates that the unit is ready to dispense the total volume (5mL) with blowout.

6.Position the pipette tip to dispense and press the Start Button. The unit will

dispense the full amount from the tip.

The display now indicates that the unit is ready to repeat the dilution sequence, beginning with the 4.50mL aspiration.

9

10

Image 5
Contents Description 3M Electronic Pipettor 5mL includes PrecautionsIndications Charging the Pipettor Battery Connecting the AC AdapterInstructions For Use Condition Range or LimitMode Selection and Mode Recall Power on Display and StartupPipettor Programming Program MenuDilution 5mL Total Volume Dilution 1mL Total Volume Repeat Dispensing 1-5 repetitions of 1mL 1mL Pipetting Open Dilution Programming 5mL PipettingSpeed Selection Er 1 Display TroubleshootingBubbles Frozen DisplayCare and Maintenance Care and MaintenanceCleaning the Tip Cone Filter ReplacementCalibration Check Chart Battery ReplacementCalibration Check Mode d1, 15, 5mL dilution Pipettor Tips SpecificationsAccuracy and Precision TolerancesLimitation of Remedies Limited WarrantyOut of Warranty Service