Pipettor Programming
5mL Pipetting
1.Press repeatedly until the display reads:
2.Press to enter selection.
3.Attach a pipette tip to the tip cone.
The pipettor is now ready to aspirate 5.00mL.
4.Place the tip in the sample and press
the Start Button. The unit will aspirate 5.00mL.
The next display indicates that the unit is ready to dispense the total volume (5mL) with blowout.
5. Position the pipette tip to dispense and
press the Start Button. The unit will dispense the full amount from the tip.
The display now indicates that the unit is ready to repeat the 5mL pipetting, beginning with the 5.00mL aspiration.
Page 9 of 15 - June 2001
Pipettor Programming
Open Dilution Programming
1.Press repeatedly until the display reads:
2.Press to enter selection. The display shows the default setting for the diluent aspiration volume.
3.Press or
to select a new diluent volume (0.10 - 5.00mL). (In this case 4.50mL)
4.Press to enter selection.
The display shows a default setting for the sample aspiration volume.
5.Press or
to select a new sample volume (In this case 0.50mL)
6.Press to enter selection.