3M 1210 instruction manual 5mL Pipetting, Open Dilution Programming

Page 9

Pipettor Programming

5mL Pipetting

1.Press repeatedly until the display reads:

2.Press to enter selection.

3.Attach a pipette tip to the tip cone.

The pipettor is now ready to aspirate 5.00mL.

4.Place the tip in the sample and press

the Start Button. The unit will aspirate 5.00mL.

The next display indicates that the unit is ready to dispense the total volume (5mL) with blowout.

5. Position the pipette tip to dispense and

press the Start Button. The unit will dispense the full amount from the tip.

The display now indicates that the unit is ready to repeat the 5mL pipetting, beginning with the 5.00mL aspiration.

Page 9 of 15 - June 2001

Pipettor Programming

Open Dilution Programming

1.Press repeatedly until the display reads:

2.Press to enter selection. The display shows the default setting for the diluent aspiration volume.

3.Press or to select a new diluent volume (0.10 - 5.00mL). (In this case 4.50mL)

4.Press to enter selection.

The display shows a default setting for the sample aspiration volume.

5.Press or to select a new sample volume (In this case 0.50mL)

6.Press to enter selection.

17

18

Image 9
Contents Description Precautions Indications3M Electronic Pipettor 5mL includes Charging the Pipettor Battery Connecting the AC AdapterInstructions For Use Condition Range or LimitMode Selection and Mode Recall Power on Display and StartupPipettor Programming Program MenuDilution 5mL Total Volume Dilution 1mL Total Volume Repeat Dispensing 1-5 repetitions of 1mL 1mL Pipetting Open Dilution Programming 5mL PipettingSpeed Selection Er 1 Display TroubleshootingBubbles Frozen DisplayCare and Maintenance Care and MaintenanceCleaning the Tip Cone Filter ReplacementBattery Replacement Calibration Check Mode d1, 15, 5mL dilutionCalibration Check Chart Pipettor Tips SpecificationsAccuracy and Precision TolerancesLimited Warranty Out of Warranty ServiceLimitation of Remedies