Pipettor Programming

The Program Menu

1. The following five dilutions have been pre-programmed.

 

Dilution

Diluent

Sample

Total

Mode

Ratio

Volume

Volume

Volume

d1

1:5

4.00mL

1.00mL

5.00mL

d2

1:10

4.50mL

0.50mL

5.00mL

d3

1:20

4.75mL

0.25mL

5.00mL

d4

1:5

0.80mL

0.20mL

1.00mL

d5

1:10

0.90mL

0.10mL

1.00mL

2. Repeat Dispense 1mL Mode-

This mode performs repeat 1mL dispensing.

Mode Volume Repetitions

rd1.00mL 1-5

3. Pipette Modes -

These modes perform standard 1mL and 5mL pipetting with blow-out.

Mode Volume

P1 1.00mL

P5 5.00mL

4. Open Dilution Mode -

This mode allows selection of any dilution ratio within the range of the pipettor

 

 

Diluent

Sample

Max.

 

 

Volume

Volume

Total

Mode

Increment

Range

Range

Volume

d0

0.05mL

0.10-5.00mL

0.00-4.90mL.

5.00mL

Page 4 of 15 - June 2001

Pipettor Programming

Power “ON” Display and Startup

The display of a switched “ON” pipettor indicates that programming has been reset to default settings.

1.Switch the unit “ON”. Display:

2.Press the Start Button twice. The display shows that the unit is ready for the default 1:5 dilution (Mode d1, 5mL) or for program changes.

Switching the pipettor “OFF” and then “ON” will reset the unit to the default settings (Mode d1, 1:5, 5mL dilution).

Mode Selection and Mode Recall

Selecting available modes on the pipettors is done with the use of these three keys:

1.Press once to display the default dilution mode of the pipettor.

2.Press repeatedly to view each of the modes available.

3.Press when the desired mode is displayed. This will enter the mode selection and allow you to use this mode.

Mode recall allows you to flip back and forth between the last two modes used.

or may not be used in the middle of an aspiration or dispensing cycle.

7

8

Image 4
Contents Description Indications Precautions3M Electronic Pipettor 5mL includes Connecting the AC Adapter Charging the Pipettor BatteryInstructions For Use Condition Range or LimitPower on Display and Startup Mode Selection and Mode RecallPipettor Programming Program MenuDilution 5mL Total Volume Dilution 1mL Total Volume Repeat Dispensing 1-5 repetitions of 1mL 1mL Pipetting 5mL Pipetting Open Dilution ProgrammingSpeed Selection Troubleshooting Er 1 DisplayBubbles Frozen DisplayCare and Maintenance Care and MaintenanceCleaning the Tip Cone Filter ReplacementCalibration Check Mode d1, 15, 5mL dilution Battery ReplacementCalibration Check Chart Specifications Pipettor TipsAccuracy and Precision TolerancesOut of Warranty Service Limited WarrantyLimitation of Remedies