JAI CV-M300 Ccir Non-interlaced / Field accumulation mode, Pulse Width Control Mode

Page 9

CV-M300

CCIR (Non-interlaced / Field accumulation mode)

Edge pre-select mode (for non-interlace only)

 

 

CCIR : 1H = 64s

 

TRIG

 

 

 

 

1H

TRIG in

 

>1H

 

 

 

 

 

EXT HD

 

 

 

 

 

 

 

INT HD

 

 

 

 

 

 

 

INT VD

 

 

 

 

 

 

 

1.5H

7.5H

 

 

7.5H

6.5H

1.5H

7.5H

EEN

 

 

 

 

 

 

 

Exposure time

 

16.5H

 

 

Exposure time

 

0.5H

0.5H

 

 

 

 

 

 

 

WEN

9H

7.5H

10H

 

14H

 

 

 

 

 

 

CCD out

 

 

 

Effective pixels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No.10 to 251

 

 

 

Video out

 

25H

 

287H

25H

 

 

 

 

 

 

 

 

 

Blanking

 

Effective video

Blanking

 

 

6.3.2. Pulse Width Control Mode

The pulse width control mode will only work in non-interlaced field accumulation mode.

The exposure is controlled by the low period of the ext. trigger pulse. The CV-M300 starts the exposure at the first HD pulse after the falling edge of the ext. trigger pulse. The exposure ends at the first HD pulse after the rising edge of the ext. trigger. The Shutter can be controlled to be within the range from >1H (>64 usec.) to <60 msec. The AC coupling causes the upper limit.

EEN (Exposure ENable) pulse and WEN (Write ENable) pulses are generated and output from the camera.

The EEN pulse indicates the exposure time. The EEN pulse is output from pin 2 of the 6-pin multi connector. The signal level is 4.0 Vp-p from a 75 Ohm source.

The WEN pulse indicates the time period of the effective video signal output and is useful for the timing and interfacing of external devices such as frame grabbers. The WEN pulse is output from pin 6 of the 6- pin multi connector. The signal level is 4.0 Vp-p from a 75 Ohm source

The video must be read out before a new trigger can be applied. The interval between trigger pulses must be longer than the time for 1 field + the shutter time. It is the limit for the field rate.

To use this mode

Set:

SW1-1, 2 and 3 to OFF

 

SW1-4 to ON for ext. trigger shutter

 

SW1-5 to OFF for field accumulation

 

SW1-6 to ON for non-interlaced

 

JP14 on PK8042 to OPEN

Input:

Ext. trigger to pin 5 on 6-pin multi connector.

 

Ext. HD to pin 6 on 12-pin multi connector. (If used).

 

75 Ohm termination is done with SW2-1 (HD) on PK8041 board.

Refer to Timing Chart and Cautions on next page.

Detailed switch and jumper setting is described in “7. Mode Setting”.

For connections see “5. Pin Assignment”.

- 8 -

Image 9
Contents CV-M300 Declaration of Conformity Table of Contents General Standard Composition Main FeaturesLocations and Functions Pin Multi-connector DC-IN/SYNC Pin Multi-connector TriggerPin Assignment Normal Shutter Mode Input/Output of HD/VD SignalExternal Trigger Mode Functions and Operations=4.4∝ S EIA Non-interlaced / Field accumulation modePulse Width Control Mode Ccir Non-interlaced / Field accumulation modePulse width control mode for non-interlace only Start/Stop Trigger Mode CV-M300 EIA Ccir Long Time Exposure ModeCV-M300 1. SW1 Switch on the Rear Panel Mode SettingTable for SW 1 Setting OFF = On =SW2 on PK8041 Board Pixel Clock Output Jumper SettingsJP10JP12 Adjustment of Video Signal Output Level Tripod mount plate External AppearanceSpectral Sensitivity SpecificationsPrecautions Typical CCD CharacteristicsAppendix User’s Record Users Mode Settings Users ModificationsCamera type CV-M300 Scanning system Revision Revision E Serial No2700-11088E E1