Hach 6120118 user manual Section Installation, Wiring the sc Sensor to the Digital Gateway

Page 11

Section 3

Installation

DANGER

Only qualified personnel should conduct the tasks described in this section of the manual.

DANGER

Seul un technicien qualifié peut effectuer les tâches d'installation décrites dans cette section du manuel.

The Combination pH/ORP Sensor can be used with either an sc100 or sc1000 controller. Refer to section 3.2 on page 13 for sc100 installation instructions and section 3.3 on page 16 for sc1000 installation instructions.

The sc sensor should be wired to the digital gateway before connecting the digital gateway to the sc100 or sc1000 Controller. The digital gateway is designed to provide a digital interface to the appropriate digital controller. Refer to section 3.1 for more information.

3.1 Wiring the sc Sensor to the Digital Gateway

DANGER

The sc100 and certain versions of the sensor are suitable for use in Class 1, Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing 58600-78 in the sc100 Controller Manual, Cat. No. 58600-18 for acceptable sensor versions and installation requirements.

DANGER

Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits dangereux de la Classe 1, Division 2, Groupes A, B, C, D. Reportez-vous au schéma de contrôle 58600-78 du Manuel du contrôleur sc100, Réf. 58600-18 pour connaître les versions des capteurs admises et les conditions d'installation.

1.Route the cable from the sensor through the strain relief in the digital gateway then properly terminate the wire ends (see Figure 5).

Note: Do not tighten the strain relief until the digital gateway is wired and the two halves are threaded securely together.

2.Insert the wires as shown in Table 3 and Figure 6.

3.Make sure the O-ring is properly installed between the two halves of the digital gateway and thread the two halves together. Hand tighten.

4.Tighten the strain relief to secure the sensor cable.

5.Connect the digital gateway to the controller.

sc100 Non-hazardous Location–section 3.2.1.1 on page 13

sc100 Hazardous Location—section 3.2.2.1 on page 16

sc1000 Connection Instructions—Refer to section 3.3 on page 16.

11

Image 11
Contents Digital PC sc and RC sc ¾-inch Combination pH/ORP Sensor Page Table of Contents Table of Contents Digital Gateway Specifications SpecificationsCombination pH and ORP Sensor Specifications Page Use of Hazard Information Safety InformationPrecautionary Labels General InformationSensor Body Styles General Sensor InformationGeneral Information Digital Gateway Operating PrecautionPage Wiring the sc Sensor to the Digital Gateway Section InstallationProper Wire Preparation and Insertion InstallationWiring the Digital Gateway Cat. No Hard-wiring the Digital Gateway to the sc100 Terminal Number Terminal Designation Wire Color Wiring the Sensor at Terminal Block J5Connecting the Sensor using the Quick-connect Fittings Connecting the Sensor to the sc1000Mounting the Digital Gateway Digital Gateway DimensionsPage Number Key Function Using the sc100 ControllerController Key Functions/Features User Interface and NavigationDisplay User Interface and Navigation Controller Display FeaturesImportant Key Presses User Interface and Navigation Using the sc1000 ControllerDisplay Features Main Menu Navigating the Menu WindowsKeypad Message window List BoxOperation Sensor SetupSensor Diagnostics Menu for pH and ORP PH Sensor Setup MenuOperation ConfigureAC Frequency ORP Sensor Setup MenuTwo Point Automatic Calibration PH CalibrationOne Point Automatic Calibration Two Point Manual Calibration Operation One Point Manual CalibrationOne-point Manual Calibration ORP CalibrationConcurrent Calibration of Two Sensors for pH and ORP Adjusting the Temperature Page Maintenance Schedule MaintenanceMaintenance Cleaning the SensorError Codes TroubleshootingError Codes Troubleshooting the pH or ORP SensorPage Catalog Number Replacement Parts and AccessoriesReplacement Items, Accessories, and Reagent and Standards Item DescriptionPage Information Required How to OrderTechnical and Customer Service U.S.A. only A. CustomersCanada Repair ServiceLimitation of Remedies Limited WarrantyLimitations Page Emissions Product SafetyCompliance Information ImmunityCompliance Information FCC Part 15, Class a LimitsGroup Name Tag Name Register Data Length Description Type Appendix a Modbus Register InformationSensor Modbus Registers Modbus Register Information Where Appendix B General pH InformationPH Measurement Theory General pH Information PID Controller BasicsDerivative IntegralPage Index