Operation
5.6.3One Point Manual Calibration
1.From the Main Menu, select SENSOR SETUP and confirm.
2.Select the appropriate sensor if more than one is attached and confirm.
3.Select CALIBRATE and confirm.
4.Select 1 POINT MANUAL. Select the available Output Mode (Active, Hold, or Transfer) from the list box and confirm.
5.Move the clean probe to solution and confirm to continue.
6.Confirm when stable. Edit the solution value using the keypad and confirm.
7.Confirm when stable. A screen will display 1 Point Manual Complete and the slope (XX.X mV/pH).
8.Return the probe to process.
5.6.4Two Point Manual Calibration
1.From the Main Menu, select SENSOR SETUP and confirm.
2.Select the appropriate sensor if more than one is attached and confirm.
3.Select CALIBRATE and confirm.
4.Select 2 POINT MANUAL CAL. Select the available Output Mode (Active, Hold, or Transfer) from the list box and confirm.
5.Move the clean probe to Solution 1 and confirm.
6.Confirm when stable. Edit the solution value using the keypad and confirm.
7.Move probe to solution 1 and confirm.
8.Confirm when stable. Edit the solution value using the keypad and confirm.
9.A screen will display 2 Point Manual Cal Complete and the slope (XX.X mV/pH).
10.Return the probe to process.
29