5.7 ORP Calibration

5.7.1 One-point Manual Calibration

The manufacturer offers a one point manual calibration for ORP. The value of the sample used in the manual calibration may be determined by laboratory analysis or comparison reading.

1.From the Main Menu, select SENSOR SETUP and confirm.

2.Select the appropriate sensor if more than one is attached and confirm.

3.Select CALIBRATE and confirm.

4.Select 1 POINT MANUAL CAL. Select the available Output Mode (Active, Hold, or Transfer) from the list box and confirm.

5.Move the clean probe to Solution and confirm.

6.Confirm when stable. Edit the solution value using the keypad and confirm.

7.A screen will display 1 Point Manual Complete and the slope (XX.X mV).

8.Return the probe to process.

5.8Concurrent Calibration of Two Sensors for pH and ORP

1.Begin a calibration on the first sensor and continue until “Wait to Stabilize” is displayed.

2.Select LEAVE and confirm. The display will return to the main measurement screen. The reading for the sensor currently being calibrated will flash.

3.Begin the calibration for the second sensor and continue until “Wait to Stabilize” is displayed.

4.Select LEAVE and confirm. The display will return to the main measurement screen and the reading for both sensors will flash. The calibration for both sensors are now running in the background.

5.To return to the calibration of either sensor select SENSOR SETUP from the Main Menu and confirm. Select the appropriate sensor and confirm.

6.The calibration in progress will be displayed. Continue with the calibration.

30

Image 30
Contents Digital PC sc and RC sc ¾-inch Combination pH/ORP Sensor Page Table of Contents Table of Contents Specifications Combination pH and ORP Sensor SpecificationsDigital Gateway Specifications Page General Information Safety InformationPrecautionary Labels Use of Hazard InformationGeneral Sensor Information General InformationSensor Body Styles Operating Precaution Digital GatewayPage Section Installation Wiring the sc Sensor to the Digital GatewayInstallation Proper Wire Preparation and InsertionWiring the Digital Gateway Cat. No Hard-wiring the Digital Gateway to the sc100 Wiring the Sensor at Terminal Block J5 Terminal Number Terminal Designation Wire ColorConnecting the Sensor to the sc1000 Connecting the Sensor using the Quick-connect FittingsDigital Gateway Dimensions Mounting the Digital GatewayPage User Interface and Navigation Using the sc100 ControllerController Key Functions/Features Number Key FunctionUser Interface and Navigation Controller Display Features Important Key PressesDisplay Using the sc1000 Controller Display FeaturesUser Interface and Navigation Navigating the Menu Windows Main MenuKeypad List Box Message windowPH Sensor Setup Menu Sensor SetupSensor Diagnostics Menu for pH and ORP OperationConfigure OperationORP Sensor Setup Menu AC FrequencyPH Calibration One Point Automatic CalibrationTwo Point Automatic Calibration Operation One Point Manual Calibration Two Point Manual CalibrationORP Calibration Concurrent Calibration of Two Sensors for pH and ORPOne-point Manual Calibration Adjusting the Temperature Page Maintenance Maintenance ScheduleCleaning the Sensor MaintenanceTroubleshooting the pH or ORP Sensor TroubleshootingError Codes Error CodesPage Item Description Replacement Parts and AccessoriesReplacement Items, Accessories, and Reagent and Standards Catalog NumberPage A. Customers How to OrderTechnical and Customer Service U.S.A. only Information RequiredRepair Service CanadaLimited Warranty LimitationsLimitation of Remedies Page Immunity Product SafetyCompliance Information EmissionsFCC Part 15, Class a Limits Compliance InformationAppendix a Modbus Register Information Sensor Modbus RegistersGroup Name Tag Name Register Data Length Description Type Modbus Register Information Appendix B General pH Information PH Measurement TheoryWhere PID Controller Basics General pH InformationIntegral DerivativePage Index