General Information

2.2 General Sensor Information

Optional equipment, such as mounting hardware for the probe, is supplied with instructions for all user installation tasks. Several mounting options are available, allowing the probe to be adapted for use in many different applications.

2.2.1 Sensor Body Styles

Combination pH and ORP sensors are available in three body styles:

Convertible Body Style (Figure 1 and Figure 2)— has ¾-in. NPT threads at both ends of the body for mounting in any of the following configurations:

into a standard ¾-in. NPT pipe tee

onto the end of a pipe for immersion into a vessel

Insertion Body Style (Figure 3)— for mounting into the pipe adapter of a ball valve hardware assembly. This hardware enables the sensor to be inserted into or retracted from the process without stopping the process flow.

Sanitary Body Style (Figure 4)— features a built-in 2-in. flange for mounting into a 2-in. sanitary tee. Included with the sanitary-style sensor is a special cap and EDPM compound gasket for use with the sanitary hardware.

Figure 1 Convertible Style Sensor with Flat Electrode

¾-inch NPT

Typical

22.2 mm (0.875 inch)

22.9mm (0.90 inch) 149.9 mm (5.9 inches)

7.62mm

(1.30 inches)

Figure 2 Convertible-style Sensor with Dome Electrode

¾-inch NPT

Typical

22.2 mm (0.875 inch)

22.9mm (0.90 inch) 149.9 mm (5.9 inches)

7.62mm

(1.30 inches)

8

Image 8
Contents Digital PC sc and RC sc ¾-inch Combination pH/ORP Sensor Page Table of Contents Table of Contents Digital Gateway Specifications SpecificationsCombination pH and ORP Sensor Specifications Page Safety Information Precautionary LabelsGeneral Information Use of Hazard InformationSensor Body Styles General Sensor InformationGeneral Information Operating Precaution Digital GatewayPage Section Installation Wiring the sc Sensor to the Digital GatewayInstallation Proper Wire Preparation and InsertionWiring the Digital Gateway Cat. No Hard-wiring the Digital Gateway to the sc100 Wiring the Sensor at Terminal Block J5 Terminal Number Terminal Designation Wire ColorConnecting the Sensor to the sc1000 Connecting the Sensor using the Quick-connect FittingsDigital Gateway Dimensions Mounting the Digital GatewayPage Using the sc100 Controller Controller Key Functions/FeaturesUser Interface and Navigation Number Key FunctionDisplay User Interface and Navigation Controller Display FeaturesImportant Key Presses User Interface and Navigation Using the sc1000 ControllerDisplay Features Navigating the Menu Windows Main MenuKeypad List Box Message windowSensor Setup Sensor Diagnostics Menu for pH and ORPPH Sensor Setup Menu OperationConfigure OperationORP Sensor Setup Menu AC FrequencyTwo Point Automatic Calibration PH CalibrationOne Point Automatic Calibration Operation One Point Manual Calibration Two Point Manual CalibrationOne-point Manual Calibration ORP CalibrationConcurrent Calibration of Two Sensors for pH and ORP Adjusting the Temperature Page Maintenance Maintenance ScheduleCleaning the Sensor MaintenanceTroubleshooting Error CodesTroubleshooting the pH or ORP Sensor Error CodesPage Replacement Parts and Accessories Replacement Items, Accessories, and Reagent and StandardsItem Description Catalog NumberPage How to Order Technical and Customer Service U.S.A. onlyA. Customers Information RequiredRepair Service CanadaLimitation of Remedies Limited WarrantyLimitations Page Product Safety Compliance InformationImmunity EmissionsFCC Part 15, Class a Limits Compliance InformationGroup Name Tag Name Register Data Length Description Type Appendix a Modbus Register InformationSensor Modbus Registers Modbus Register Information Where Appendix B General pH InformationPH Measurement Theory PID Controller Basics General pH InformationIntegral DerivativePage Index