Page 16

Installation

3.2.2.1 Attaching the Digital Gateway with a Quick-Connect Fitting in a Hazardous Location

The digital gateway is supplied with a keyed quick-connect fitting for easy attachment to the controller, see Figure 7 on page 14. For Hazardous Locations, a connector safety lock (Cat. No. 6139900) must be installed. Retain the connector cap to seal the connector opening in case the sensor must be removed.

1.Remove the connector cap from the sc100 controller. Retain the connector cap to seal the connector opening in case the sensor must be removed.

2.Connect the digital gateway connector to the sensor plug on the sc100.

3.Install the connector safety lock (Figure 9). Align the lock over the connector and squeeze the two halves together to lock. To remove the connector safety lock, insert a small flat-blade screwdriver into the locking groove. Pivot the screwdriver away from the groove and separate the two halves (Figure 9).

Figure 9 Installing the Connector Safety Lock

38.1 mm

(1.50 inches)

38.1 mm

(1.50 inches)

3.3 Connecting the Sensor to the sc1000

3.3.1Connecting the Sensor using the Quick-connect Fittings

1.Unscrew the connector cap from the controller. Retain the connector cap to seal the connector opening in case the sensor must be removed.

2.Push the digital gateway connector into the socket.

3.Hand-tighten the union nut.

Note: Do not use the middle connection for the sensors as this is reserved for the display module.

16

Image 16
Contents Digital PC sc and RC sc ¾-inch Combination pH/ORP Sensor Page Table of Contents Table of Contents Combination pH and ORP Sensor Specifications SpecificationsDigital Gateway Specifications Page Safety Information Precautionary LabelsGeneral Information Use of Hazard InformationGeneral Information General Sensor InformationSensor Body Styles Operating Precaution Digital GatewayPage Section Installation Wiring the sc Sensor to the Digital GatewayInstallation Proper Wire Preparation and InsertionWiring the Digital Gateway Cat. No Hard-wiring the Digital Gateway to the sc100 Wiring the Sensor at Terminal Block J5 Terminal Number Terminal Designation Wire ColorConnecting the Sensor to the sc1000 Connecting the Sensor using the Quick-connect FittingsDigital Gateway Dimensions Mounting the Digital GatewayPage Using the sc100 Controller Controller Key Functions/FeaturesUser Interface and Navigation Number Key FunctionImportant Key Presses User Interface and Navigation Controller Display FeaturesDisplay Display Features Using the sc1000 ControllerUser Interface and Navigation Navigating the Menu Windows Main MenuKeypad List Box Message windowSensor Setup Sensor Diagnostics Menu for pH and ORPPH Sensor Setup Menu OperationConfigure OperationORP Sensor Setup Menu AC FrequencyOne Point Automatic Calibration PH CalibrationTwo Point Automatic Calibration Operation One Point Manual Calibration Two Point Manual CalibrationConcurrent Calibration of Two Sensors for pH and ORP ORP CalibrationOne-point Manual Calibration Adjusting the Temperature Page Maintenance Maintenance ScheduleCleaning the Sensor MaintenanceTroubleshooting Error CodesTroubleshooting the pH or ORP Sensor Error CodesPage Replacement Parts and Accessories Replacement Items, Accessories, and Reagent and StandardsItem Description Catalog NumberPage How to Order Technical and Customer Service U.S.A. onlyA. Customers Information RequiredRepair Service CanadaLimitations Limited WarrantyLimitation of Remedies Page Product Safety Compliance InformationImmunity EmissionsFCC Part 15, Class a Limits Compliance InformationSensor Modbus Registers Appendix a Modbus Register InformationGroup Name Tag Name Register Data Length Description Type Modbus Register Information PH Measurement Theory Appendix B General pH InformationWhere PID Controller Basics General pH InformationIntegral DerivativePage Index