4
WARNING
The
Electrical Specifications HF185, IHF185, IHF045, CHF185, CHF045
185 liters / 45 pounds
|
|
|
|
| Boiler Fed/Steam Heat |
| Electric Heat |
| ||||||
Code | Voltage | Cycle | Phase | Wire | Full Load Amps |
| Recommended Circuit Breaker | AWG/mm2 | kW Standard Heating Elements | Full Load Amps |
| Recommended Circuit Breaker (US- market) | AWG/mm2 | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| US |
|
|
| US |
|
| ||
N | 50/60 | 3 | 3+PE | 6 | 15 | 10 | 14/2.5 |
| 32 | 40 |
| 40 | 8/10.0 | |
P | 50/60 | 3 | 3+N+PE | 18 | 20 | 20 | 12/4.0 | 6x3 kW | 44 | 50 |
| 50 | 8/10.0 | |
Q | 50/60 | 3 | 3+PE | 18 | 20 | 20 | 12/4.0 | 63 | 70 |
| 70 | 4/25.0 | ||
|
| |||||||||||||
X | 50/60 | 1/3 | 2/3+PE | 18 | 20 | 20 | 12/4.0 |
| N/A | N/A |
| N/A | N/A | |
|
|
|
|
|
|
|
|
| Alternative Electric Heat Options | |||||
N | 50/60 | 3 | 3+PE |
|
|
|
|
| 23 | 30 |
| 32 | 10/6.0 | |
P | 50/60 | 3 | 3+N+PE |
|
|
|
| 6x2 kW | 36 | 40 |
| 40 | 8/10.0 | |
Q | 50/60 | 3 | 3+PE |
|
|
|
| 49 | 60 |
| 60 | 6/16.0 | ||
|
|
|
|
|
| |||||||||
X | 50/60 | 1/3 | 2/3+PE |
|
|
|
|
| N/A | N/A |
| N/A | N/A | |
N | 50/60 | 3 | 3+PE |
|
|
|
|
| 28 | 40 |
| 40 | 8/10.0 | |
P | 50/60 | 3 | 3+N+PE |
|
|
|
| 3x3 kW + | 40 | 50 |
| 50 | 8/10.0 | |
Q | 50/60 | 3 | 3+PE |
|
|
|
| 3x2 kW | 56 | 60 |
| 60 | 6/16.0 | |
X | 50/60 | 1/3 | 2/3+PE |
|
|
|
|
| N/A | N/A |
| N/A | N/A | |
N | 50/60 | 3 | 3+PE |
|
|
|
|
| 36 | 40 |
| 40 | 8/10.0 | |
P | 50/60 | 3 | 3+N+PE |
|
|
|
| 3x3 kW + | 49 | 60 |
| 60 | 6/16.0 | |
Q | 50/60 | 3 | 3+PE |
|
|
|
| 3x4 kW | N/A | N/A |
| N/A | N/A | |
X | 50/60 | 1/3 | 2/3+PE |
|
|
|
|
| N/A | N/A |
| N/A | N/A | |
N | 50/60 | 3 | 3+PE |
|
|
|
|
| 41 | 50 |
| 50 | 8/10.0 | |
P | 50/60 | 3 | 3+N+PE |
|
|
|
| 6x4 kW | 53 | 60 |
| 60 | 6/16.0 | |
Q | 50/60 | 3 | 3+PE |
|
|
|
| N/A | N/A |
| N/A | N/A | ||
|
|
|
|
|
| |||||||||
X | 50/60 | 1/3 | 2/3+PE |
|
|
|
|
| N/A | N/A |
| N/A | N/A | |
|
|
|
|
|
|
| Table 4 |
|
|
|
|
|
|
|
32