Dakota Digital MMX-6 operation manual Calibration

Page 19

MMX-6 DL Ultrasonic Thickness Gauge

6)While the transducer is firmly coupled to the probe-disc, press the PRB-0key on the keypad. The MMX-6 DL will display "Prb0" while it is calculating its zero point.

7)Remove the transducer from the probe-disc.

At this point, the MMX-6 DL has successfully calculated it's internal error factor, and will compensate for this value in any subsequent measurements. When performing a "probe-zero", the MMX-6 DL will always use the sound-velocity value of the built-in probe-disc, even if some other velocity value has been entered for making actual measurements. Though the MMX-6 DL will remember the last "probe-zero" performed, it is generally a good idea to perform a "probe-zero" whenever the tool is turned on, as well as any time a different transducer is used. This will ensure that the instrument is always correctly zeroed.

Calibration

In order for the MMX-6 DL to make accurate measurements, it must be set to the correct sound-velocity for the material being measured. Different types of material have different inherent sound-velocities. For example, the velocity of sound through steel is about 0.233 inches-per-microsecond, versus that of aluminum, which is about 0.248 inches-per-microsecond. If the tool is not set to the correct sound-velocity, all of the measurements the tool makes will be erroneous by some fixed percentage. The one point

15

Image 19
Contents Dakota Ultrasonics Dakota Ultrasonics Contents Disclaimer Very Important Introduction Blank Keypad OperationDakota Ultrasonics MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Display Dakota Ultrasonics Is currently operating in echo-echo Thru-paint/coating mode Transducer Making Measurements Dakota Ultrasonics Condition and Preparation of Surfaces Performing a Probe-Zero Probe ZeroCalibration Dakota Ultrasonics Calibration to a known thickness Dakota Ultrasonics Calibration to a known velocity Dakota Ultrasonics Two Point Calibration Units in/mm MMX-6 DL Modes Changing Units in / MMScan Mode Using The Back LightBack light Alarm Mode Using Scan ModeUsing the Beeper Dual-Multi Mode RS232 Serial Port & The Data Logger Connecting To a ComputerUsing The Data Logger Using the Data Logger Clearing a Storage Location Clearing an Entire File Sending All Files to a Computer Clearing all FilesPrinting a File Transducer Selection MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Blank Keypad Product Specifications PhysicalPower Source Appendix aApplication Notes Measuring pipe and tubing Appendix BMeasuring hot surfaces Measuring laminated materials Measuring Through Paint & Coatings Blank Sound Velocities of some Common Materials Appendix CBlank Warranty Statement Warranty Information