varies from lot to lot and from manufacturer to manufacturer. Calibration to a sample of known thickness will ensure that the tool is set as closely as possible to the sound velocity of the material to be measured.

20

Image 24
Contents Dakota Ultrasonics Dakota Ultrasonics Contents Disclaimer Very Important Introduction Blank Operation KeypadDakota Ultrasonics MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Display Dakota Ultrasonics Is currently operating in echo-echo Thru-paint/coating mode Transducer Making Measurements Dakota Ultrasonics Condition and Preparation of Surfaces Probe Zero Performing a Probe-ZeroCalibration Dakota Ultrasonics Calibration to a known thickness Dakota Ultrasonics Calibration to a known velocity Dakota Ultrasonics Two Point Calibration MMX-6 DL Modes Changing Units in / MM Units in/mmUsing The Back Light Scan ModeBack light Using Scan Mode Alarm ModeUsing the Beeper Dual-Multi Mode Connecting To a Computer RS232 Serial Port & The Data LoggerUsing The Data Logger Using the Data Logger Clearing a Storage Location Clearing an Entire File Clearing all Files Sending All Files to a ComputerPrinting a File Transducer Selection MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Blank Product Specifications Physical Power SourceAppendix a KeypadAppendix B Application Notes Measuring pipe and tubingMeasuring hot surfaces Measuring laminated materials Measuring Through Paint & Coatings Blank Appendix C Sound Velocities of some Common MaterialsBlank Warranty Information Warranty Statement