Dakota Digital MMX-6 operation manual Clearing an Entire File

Page 35

MMX-6 DL Ultrasonic Thickness Gauge

Clearing an Entire File

The user may require the contents of an entire file be completely cleared of all measurements. This would allow the user to start a new list of measurements starting at storage location L001. The procedure is outlined in the following steps:

1)Press the ON/OFF key to power up the MMX-6 DL.

2)Press the MEM key to activate the data logging functions and settings.

3)Press the SEND key to enter file setup.

4)Press the UP / DOWN arrow keys to scroll to the file that will be cleared of all measurements.

5)Press the SEND key once again to select the file.

Note:

ƒThe display will flash the FILE / F-05( The file selected by the user ) symbol.

6)Press the UP / DOWN arrow keys to scroll to the flashing CLr / F- 05 ( The file selected by the user ) symbol.

7)Press the SEND key to select the clear file option. The symbol ( CLr? ) will be displayed.

8)Press the CLR key to confirm and clear the contents of the entire file.

9)Press the MEM key, at any time, to exit the data logging functions and return to measurement mode.

31

Image 35
Contents Dakota Ultrasonics Dakota Ultrasonics Contents Disclaimer Very Important Introduction Blank Keypad OperationDakota Ultrasonics MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Display Dakota Ultrasonics Is currently operating in echo-echo Thru-paint/coating mode Transducer Making Measurements Dakota Ultrasonics Condition and Preparation of Surfaces Performing a Probe-Zero Probe ZeroCalibration Dakota Ultrasonics Calibration to a known thickness Dakota Ultrasonics Calibration to a known velocity Dakota Ultrasonics Two Point Calibration Units in/mm MMX-6 DL Modes Changing Units in / MMBack light Using The Back LightScan Mode Alarm Mode Using Scan ModeUsing the Beeper Dual-Multi Mode RS232 Serial Port & The Data Logger Connecting To a ComputerUsing The Data Logger Using the Data Logger Clearing a Storage Location Clearing an Entire File Sending All Files to a Computer Clearing all FilesPrinting a File Transducer Selection MMX-6 DL Ultrasonic Thickness Gauge Dakota Ultrasonics Blank Keypad Product Specifications PhysicalPower Source Appendix aMeasuring hot surfaces Appendix BApplication Notes Measuring pipe and tubing Measuring laminated materials Measuring Through Paint & Coatings Blank Sound Velocities of some Common Materials Appendix CBlank Warranty Statement Warranty Information