Chapter 4 - Repairs and Replacements

Nebulizer Extension

P/N 754061

Nebulizer Cotter Pin

P/N 853071

Gauge Assembly

P/N 854117

The nebulizer extension is a 2 in. long stainless steel tube that provides a channel for the oxygen from the nebulizer to the interior of the therapy chamber (Figure 18). To mount a new extension to a new nebulizer cap, use a high quality silicone cement. Make sure the finished joint is secure and free of air leaks.

The cotter pin holds the nebulizer to the oxygen therapy door (Figure 18). If you should lose or damage this pin, we recommend that you obtain a replacement locally to save time and expense. The pin is stainless steel, 1/16 in. diameter, and 1.5 in. long. In an emergency, use a bent paper clip as a temporary replacement!

Overview

The original gauge was a square, analog-type instrument. This gauge is no longer available, and has been replaced by a smaller, more modern, digital instrument (Figure 24).

Tools Required

„3/8 in. wrench

„Phillips screwdriver

„Small flat-blade screwdriver (for old-style gauge only)

„Electric drill with 1/16 in. bit

Removal - Old-style Gauge

1.Open the cage door.

2.With a Phillips screwdriver and a 3/8 in. wrench, remove the two gauge bracket screws/nuts (Figure 22).

3.With a small flat-blade screwdriver, unscrew the two gauge screws (Figure 23).

Gauge Screws

Gauge Bracket Screws

Gauge

Bracket

Figure 22. Removing the Old-style

Figure 23.

Rear the Old-style Gauge

Gauge Assembly

Assembly

 

22

SSCI - Suburban Surgical Company, Inc.

Image 26
Contents Ssci Page Table of Contents Repairs and Replacements TroubleshootingGeneral Information Introduction About this ManualModels Oxygen Therapy Door Information Safety NoticesModels Care and Cleaning of Stainless Steel Introduction Cleaning CleansersSafety Cleaning RequirementsSsci Contact Information Warranty Telephone IllinoisComments Tools Required Installation and SetupUnpacking Inspection Parts Included Installation Setup On a Regal Cage ProcedureHinges Disposition of the Shipping Carton NebulizerComments Door and Latch Operating and CleaningOpening and Closing Latch Adjustment Door SealsGeneral Ice TrayGauges Changing Between C and F Displays Maximum/Minimum Temperature Humidity RecordingReplacing the Gauge Battery Using the Nebulizer Gauge Battery Slide and BatteryOxygen Concentration Control Oxygen Concentration ControlIntroduction Cleaning InstructionsNebulizer Cleaning Procedures Replacement Parts Available for the Oxygen Therapy Door Repairs and ReplacementsReplacement Parts General Information Parts Ordering Procedure Procedures Tool RequiredNebulizer Oxygen Supply Hose Cotter PinNebulizer Hose Barb Fitting Tools RequiredRemoval Old-style Gauge Nebulizer Extension Nebulizer Cotter Pin Gauge AssemblyOverview Removal New-style Gauge Front View of the New-style Gauge mounted on doorDrilling Button Access Holes Ice Tray Ice Tray ValveParts Identification Seal Strip Retainer and Seal StripRefer to Tables 3 and 4 for Part Numbers Retainer Mounting Screws Retainer and Seal StripDoor Model Door Dimensions Seal Strip Retainer Quantity Seal Strip Retainer Part NumbersDoor Model Door Dimensions Seal Strip P/N Quantity Seal Strip Part NumbersTroubleshooting GeneralReturning the Oxygen Therapy Door for Repairs RMA Numbers Packing and ShipmentThere is no oxygen flow into the cage Remedial ActionThere is excessive oxygen leakage from the cage Door does not close or latch correctly One or both of the gauges does not read correctly Inside back cover This page left blank intentionally Toll Free 800 Suburban Surgical Company, IncTwelfth Street Wheeling, IL Phone 847