How to Replace the Mesh Cap
You can usually use the Mesh Cap for approximately 1 year (according to the test results of OMRON). However, this may vary depending on medication type and individual use and care. In some cases, the unit may not nebulize or may nebulize weakly even after disinfection is done.
If the unit does not nebulize or nebulizes poorly even after cleaning the Mesh Cap,
purchase a new Mesh Cap from your nearest Omron dealer or where you purchased your unit from. The part number is
1Open the Mesh Cap.
Lift up the Mesh Cap by strongly applying your thumb on the brim.
2Remove the Mesh Cap from the
Medication Bottle. | Lift up the |
| |
| Mesh Cap |
| vertically. |
3Install a new Mesh Cap.
Insert a new |
| |
Mesh Cap | Close it | |
vertically. | ||
securely. | ||
|
How to Disinfect How to Replace the Mesh Cap
19