Emerson NGA 2000 TO2 instruction manual Basic CONTROLS, Setup and Status, Analyzer Channel Status

Page 32

Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

3-4 BASIC CONTROLS, SETUP AND STATUS

The following sections describe the basic control of the analyzer and the viewing of channel parameters. Examples of stepping through the menus are shown so that the user can become familiar with the operation,

a. Analyzer Channel Status

keeping in mind that displays and menu choices may be different depending on actual analyzer configuration and any customization of the menus.

TO2

 

 

 

 

 

 

 

 

 

 

 

 

 

37.50

 

 

 

 

 

 

 

 

ppm O2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.00

 

 

 

 

 

Range: 3

 

 

50.00

 

 

Electrolyte level:

 

OK

 

 

 

 

 

 

 

 

 

 

Temp current:

 

263 uA

0.0

 

 

 

100.0

 

 

Range:

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Sensor Current:

0.0500 uA

0.0000

 

 

100.0000

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display

 

 

Status…

 

Main…

 

 

Channel

 

BasicCal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TO2

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

-- TO2 Analyzer Module Status --

 

 

 

 

 

Electrolyte level:

 

 

 

 

 

 

 

 

 

Ok

 

 

Range and calibration control:

 

 

 

 

 

 

Local

 

 

Next quick start possible after:

 

 

 

 

 

 

 

0 s

 

 

Operation status:

 

 

 

 

 

 

 

 

Normal

 

 

Sensor current:

 

 

 

 

 

 

 

0.0500 uA

 

 

Sensor temperature:

 

 

 

 

 

 

 

 

25.3 °C

 

 

Sensor model:

 

 

 

 

 

 

0-100 Standard

 

 

Over range:

 

 

 

 

 

 

 

 

 

 

 

No

 

 

Power supply voltages…

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measure

 

 

Back…

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TO2

 

 

 

 

 

 

 

 

 

37.50 ppm

 

 

 

 

 

 

 

-- Power Supply Voltages --

 

 

 

 

 

+15V analog:

 

 

 

 

 

 

 

 

14.9V

 

 

-15V analog:

 

 

 

 

 

 

 

 

-15.0V

 

 

+5V digital:

 

 

 

 

 

 

 

 

 

 

5.02V

 

 

+24V power:

 

 

 

 

 

 

 

 

23.89V

 

 

Primary electrode:

 

 

 

 

 

 

 

 

1.31V

 

 

Isolated 15V:

 

 

 

 

 

 

 

 

 

OK

 

If necessary, from the Main menu press F1 “Measure.”

Press F2 to change to the “Analyzer Channel Status” menu.

Or, from the Main menu press F2 to the “Analyzer Channel Status” menu.

From the “Analyzer Module Status” menu an additional submenu is available for “Power supply voltages..." to check the various volt- age levels in the analyzer.

Press the or key to access the “Power supply voltages..." display screen. The toler- ance for these voltage is ±10% of nominal.

Press the F4 (Back…) button to return to the “Analyzer Module Status” menu or the F1 (Measure) button to return to the “Measure” display screen.

From the “Analyzer Module Status” menu, press the F2 (Back…) button to go to the “Main Menu” or press the F1 (Measure) but- ton to return to the “Measure” display screen.

Measure

Back…

Figure 3-10. Analyzer Channel Status

3-8 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 32
Contents Model NGA 2000 TO2 Essential Instructions Table of Contents Model NGA 2000 TO2System & Network I/O Module Controls Setup Material Safety Data SheetList of Illustrations Analyzer Module Interconnection with Instrument PlatformModel NGA 2000 TO2 Preface Intended USE Statement DefinitionsSafety Summary Model NGA 2000 TO2 Documentation CompliancesGlossary of Terms Softkeys SystemSection Description and Specifications FeaturesOverview Typical ApplicationsTrace Oxygen Detector Coulometric Principle Trace Oxygen Analyzer Module Top View SPECIFICATIONS1 GeneralSample PhysicalGas Connections Model NGA 2000 TO2 Section Installation UnpackingElectrolyte Electrolyte AdditionGases GAS ConnectionsGAS Overpressure Sample FlowElectrical Connections Electrical Shock Hazard Leak TestSensor Damage Analyzer Module Interconnection with Instrument Platform Outline and Mounting Dimensions Back Panel Connections Operation Displays & Operating KeysMeasure Status… Channel Lock… MFG Data Analyzer basic controls calibration & setup…Menu Lines & Softkey Functionality Measure Back… Signal on mini-bargraphEntering & Changing Variables Display Status… Main… Channel BasicCal Measure Mode DisplayPpm O2 Module Manufacturing Data Displays LCDReset Abort Startup & InitializationMeasure Status… Back… Quick StartInitiate quick start Basic CONTROLS, Setup and Status Power supply voltages… Measure Back…Analyzer Channel Status Display Status… Main… ChannelMain… Channel BasicCal Single Component DisplayDisplay Status… Select Status… Tags Off LCDReset Multi Component DisplayBasic Controls and Setup Range numberOperation Status Range numberDisplay Controls Brightness and ContrastDisplay controls… Brightness 70 %Expert Controls and Setup Analyzer and I/O, expert controls & setup…Analyzer module controls… Measure Channel Back…Calibration Procedure Gas Scale FactorGas scale factor… Air background gas concentrationSpan Calibration Span gas 90.0 ppm Measure Status… Back…Alarms Alarms…Concentration alarm setup… Measure Back… ClrAlm Alarm generation isReRanges Ranges…Begin / end of ranges… Range Begin PpmLinearization Measurement Display ConfigurationMeasurement display configuration… Concentration Measurement Gas measurement unitsConcentration measurement… Gas concentration unitsTemperature compensation at span gas Sensor modelClamp low PVA to Ppm to mg/Nm3 conversion factorFlow lower limit 400 ml/min Flow MeasurementFlow measurement… Temperature measurement units Temperature MeasurementTemperature measurement… Measure Status… Back… System & Network I/O Module CON Trols Setup System & network I/O module controls…37.50 ppm System SIO module…Analog output setup… System SIOAnalog Output Setup Choose signalSelect analog output number Choose the analyzer moduleHold output during calibration Signal value for 0% outputHold output during calibration Output number Operation modeOutputs value on analyzer failure Analog Output Updates Per Second Type of installed serial inter- faceOutput #1 Yes Baud rate 19200Relay Outputs Setup Invert signalCommunication protocol Choose source moduleChoose signal… System DIOFunction control Input numberSystem Configuration and DIAG- Nostics Analyzer Module Diagnostics Diagnostic MenusLoad/Save Module Configura- tion Date and Time Security CodesEntering a level locked by security code System ResetSystem reset Measure Back… Model NGA 2000 TO2 Sensor Caustic Liquid Section Maintenance and ServiceOverview Electrical Shock Hazard Sensor Assembly Refilling SensorSensor Lleak Check Electrolyte ReplacementSensor Replacement Analyzer and I/O expert controls & setup… LinearizationDo not tilt Sensor excessively as elec- trolyte may leak Time & DateLinearization temperature values… Linearization…0000 ppm Linearization raw values sensor current…Flow Sensor Replacement Electronic ComponentsFuses Printed Circuit BoardsSection Troubleshooting TroubleshootingFailure to purge down to ppm levels Erratic and very insensitive readingsModel NGA 2000 TO2 TO2 NGA 2000 TO2 Trace Oxygen Analyzer Module Section Replacement PartsMatrix Replacement Parts Customer Service Section Return of MaterialReturn of Material TrainingModel NGA 2000 TO2 Section Index Model NGA 2000 TO2 Product ELECTROLYTE1 Emergency & first aid procedures Precautions to be taken in handlingOther precautions Threshold limit value Routes of entryRosemount Warranty Asia Pacific EUROPE, Middle EAST, Africa Latin America