Emerson NGA 2000 TO2 Analog Output Updates Per Second, Type of installed serial inter- face

Page 53

Model NGA 2000 TO2

Instruction Manual

760008-A September 2001

Special Scaling for Concentra-

c. Serial Interface Setup

tion Signal

 

TO2

37.50 ppm

-- Special Scaling for Concentration Signal --

 

(Scaling is the same as range limits)

 

 

Output #1:

Yes

Output #2:

Yes

Output #3:

Yes

Output #4:

Yes

Output #5:

Yes

Output #6:

Yes

Output #7:

Yes

Output #8:

Yes

Measure

Back…

More…

From menu System SIO Module (Figure 3-25), select Serial interface setup…

TO2

-- Serial Interface Setup --

37.50 ppm

 

 

 

 

 

 

 

 

 

 

 

 

Baud rate:

 

 

19200

 

 

Data bits:

 

 

8

 

 

Stop bits:

 

 

1

 

 

Parity:

 

 

None

 

Echo mode:

 

 

Disabled

 

Handshake:

 

 

Xon/Xoff

 

Transmission delay:

 

0

 

 

Type of installed serial interface:

 

RS232

 

Communication protocol:

 

AK

 

Special protocol definitions…

 

 

 

 

Measure

 

 

Back…

 

 

This menu allows for the setting of each of the 8 outputs to be the same as the range limit “Yes” or as set on the previous menus.

See Section 3-5c, Ranges for setting the range limits.

Pressing F5 (More…) changes to the submenu Analog Output Updates per Second.

Pressing the F5 (More…) key changes to the submenu “Analog Output Updates per Second.”

Analog Output Updates Per

Second

TO2

 

37.50 ppm

 

 

 

-- Analog Output Updates per Second --

 

 

 

 

 

 

Output #1:

10

 

 

Output #2:

0

 

 

Output #3:

0

 

 

Output #4:

0

 

 

Output #5:

0

 

 

Output #6:

0

 

 

Output #7:

0

 

 

Output #8:

0

 

 

Measure

 

Back…

 

 

This menu allows for the setting of the update rate for each of the 8 outputs.

In this menu, the parameters for data transfer between the analyzer and exter- nal devices are set. The choices in this menu depend on the configuration of the analyzer. The full specification of the se- rial interface is described in its own man- ual.

Options:

Baud rate: 300, 1200, 2400, 4800, 9600, 19200

Data bits: 7, 8

Stop bits: 1, 2

Parity: None, Even, Odd

Echo mode: Enabled, Disabled

Handshake: None, Xon/Xoff

Transmission delay: 0…100.

Type of installed serial inter- face:

RS232

RS485/2w

RS485/4w

RS485/4w bus

None

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-29

Image 53
Contents Model NGA 2000 TO2 Essential Instructions Model NGA 2000 TO2 Table of ContentsMaterial Safety Data Sheet System & Network I/O Module Controls SetupAnalyzer Module Interconnection with Instrument Platform List of IllustrationsModel NGA 2000 TO2 Definitions Preface Intended USE StatementSafety Summary Model NGA 2000 TO2 Compliances DocumentationGlossary of Terms System SoftkeysFeatures Section Description and SpecificationsOverview Typical ApplicationsTrace Oxygen Detector Coulometric Principle Trace Oxygen Analyzer Module Top View General SPECIFICATIONS1Sample PhysicalGas Connections Model NGA 2000 TO2 Unpacking Section InstallationElectrolyte Electrolyte AdditionGAS Connections GasesGAS Overpressure Sample FlowElectrical Connections Electrical Shock Hazard Leak TestSensor Damage Analyzer Module Interconnection with Instrument Platform Outline and Mounting Dimensions Back Panel Connections Displays & Operating Keys OperationMeasure Status… Channel Lock… MFG Data Analyzer basic controls calibration & setup…Menu Lines & Softkey Functionality Measure Back… Signal on mini-bargraphEntering & Changing Variables Display Status… Main… Channel BasicCal Measure Mode DisplayPpm O2 Module Manufacturing Data Displays Startup & Initialization LCDReset AbortMeasure Status… Back… Quick StartInitiate quick start Power supply voltages… Measure Back… Basic CONTROLS, Setup and StatusAnalyzer Channel Status Display Status… Main… ChannelMain… Channel BasicCal Single Component DisplayDisplay Status… Multi Component Display Select Status… Tags Off LCDResetRange number Basic Controls and SetupOperation Status Range numberBrightness and Contrast Display ControlsDisplay controls… Brightness 70 %Analyzer and I/O, expert controls & setup… Expert Controls and SetupAnalyzer module controls… Measure Channel Back…Gas Scale Factor Calibration ProcedureAir background gas concentration Gas scale factor…Span gas 90.0 ppm Measure Status… Back… Span CalibrationAlarms… AlarmsConcentration alarm setup… Measure Back… ClrAlm Alarm generation isReRanges… RangesBegin / end of ranges… Range Begin PpmLinearization Measurement Display ConfigurationMeasurement display configuration… Gas measurement units Concentration MeasurementConcentration measurement… Gas concentration unitsSensor model Temperature compensation at span gasClamp low PVA to Ppm to mg/Nm3 conversion factorFlow lower limit 400 ml/min Flow MeasurementFlow measurement… Temperature measurement units Temperature MeasurementTemperature measurement… Measure Status… Back… System & network I/O module controls… System & Network I/O Module CON Trols Setup37.50 ppm System SIO module…System SIO Analog output setup…Choose signal Analog Output SetupSelect analog output number Choose the analyzer moduleHold output during calibration Signal value for 0% outputHold output during calibration Output number Operation modeOutputs value on analyzer failure Type of installed serial inter- face Analog Output Updates Per SecondOutput #1 Yes Baud rate 19200Invert signal Relay Outputs SetupCommunication protocol Choose source moduleSystem DIO Choose signal…Function control Input numberSystem Configuration and DIAG- Nostics Analyzer Module Diagnostics Diagnostic MenusLoad/Save Module Configura- tion Security Codes Date and TimeEntering a level locked by security code System ResetSystem reset Measure Back… Model NGA 2000 TO2 Sensor Caustic Liquid Section Maintenance and ServiceOverview Electrical Shock Hazard Refilling Sensor Sensor AssemblySensor Lleak Check Electrolyte ReplacementSensor Replacement Linearization Analyzer and I/O expert controls & setup…Do not tilt Sensor excessively as elec- trolyte may leak Time & DateLinearization… Linearization temperature values…0000 ppm Linearization raw values sensor current…Electronic Components Flow Sensor ReplacementFuses Printed Circuit BoardsTroubleshooting Section TroubleshootingFailure to purge down to ppm levels Erratic and very insensitive readingsModel NGA 2000 TO2 TO2 NGA 2000 TO2 Trace Oxygen Analyzer Module Section Replacement PartsMatrix Replacement Parts Section Return of Material Customer ServiceReturn of Material TrainingModel NGA 2000 TO2 Section Index Model NGA 2000 TO2 Product ELECTROLYTE1 Precautions to be taken in handling Emergency & first aid proceduresOther precautions Threshold limit value Routes of entryRosemount Warranty Asia Pacific EUROPE, Middle EAST, Africa Latin America