Emerson NGA 2000 TO2 Outputs value on analyzer failure, Output number, Operation mode

Page 52

Instruction Manual

760008-A September 2001

Model NGA 2000 TO2

Output(s) value on analyzer failure

TO2

37.50 ppm

-- Output Signal if Assigned Module Fails --

Output(s) value on analyzer failure:

BeginOfRange – 10%

-- Fine Adjustment –

Output number:

1

Operation mode:

Normal

Fine adjustment for 0% output:

4096

Fine adjustment for 100% output:

819

Measure

Back…

More…

Choose the desired signal level to cause a failure condition. The choices are:

Actual

Normal: The absolute measurement sig- nal will be sent to the analog output.

Adjust 0V: Used to set the display equal to the analog output for 0V and 0 mA. Life zero signals (4-20 mA and 2-10V) are set automatically and cannot be adjusted.

Adjust 10V: Used to set the display equal to the analog output for 10V and 20 mA.

Select the “Fine adjustment for 0% out- put” and/or “Fine adjustment for 100% output” lines with the or key. Adjust to the desired value with the or key and confirm with the key. The range of values are:

BeginOfRange

EndOfRange

BeginOfRange-10%

BeginOfRange+10%

Output number

TO2

 

37.50 ppm

 

-- Output Signal if Assigned Module Fails --

 

Output(s) value on analyzer failure:

BeginOfRange – 10%

 

 

-- Fine Adjustment –

 

 

 

 

Output number:

 

1

 

 

Operation mode:

 

 

Normal

 

Fine adjustment for 0% output:

 

4096

 

 

Fine adjustment for 100% output:

 

819

 

 

 

 

 

 

 

 

 

Measure

 

Back…

 

More…

Choose the output number (1-8) for set- ting the fine adjustment.

Operation mode

TO2

 

37.50 ppm

 

-- Output Signal if Assigned Module Fails --

 

Output(s) value on analyzer failure:

BeginOfRange – 10%

 

 

-- Fine Adjustment –

 

 

 

 

Output number:

 

1

 

 

Operation mode:

 

 

Normal

 

 

Fine adjustment for 0% output:

 

4096

 

 

Fine adjustment for 100% output:

 

819

 

 

 

 

 

 

 

 

 

Measure

 

Back…

 

More…

3000 to 6000 for 0% (default 4096)

600 to 1000 for 100% (default 819)

The last three lines of the “Analog Output Setup” menu are display only for configu- ration values of the analog output.

Signal name: The name of the sig- nal chosen in the “Choose signal” menu.

Current signal value: The current value of the variable.

Source module: The name of the module chosen in the “Choose sig- nal source module” menu.Pressing the F5 (More…) key changes to the submenu “Special Scaling for Con- centration Signal.”

This menu allows for the setting of each of the 8 outputs to be the same as the range limits “Yes” or as set on the previ- ous menus.

See section 3-5c for setting the range lim- its.

Pressing F5 (More…) changes to the submenu “Special Scaling for Concentra- tion Signal”.

3-28 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 52
Contents Model NGA 2000 TO2 Essential Instructions Table of Contents Model NGA 2000 TO2System & Network I/O Module Controls Setup Material Safety Data SheetList of Illustrations Analyzer Module Interconnection with Instrument PlatformModel NGA 2000 TO2 Preface Intended USE Statement DefinitionsSafety Summary Model NGA 2000 TO2 Documentation CompliancesGlossary of Terms Softkeys SystemSection Description and Specifications FeaturesOverview Typical ApplicationsTrace Oxygen Detector Coulometric Principle Trace Oxygen Analyzer Module Top View SPECIFICATIONS1 GeneralSample PhysicalGas Connections Model NGA 2000 TO2 Section Installation UnpackingElectrolyte Electrolyte AdditionGases GAS ConnectionsGAS Overpressure Sample FlowSensor Damage Leak TestElectrical Connections Electrical Shock Hazard Analyzer Module Interconnection with Instrument Platform Outline and Mounting Dimensions Back Panel Connections Operation Displays & Operating KeysMenu Lines & Softkey Functionality Analyzer basic controls calibration & setup…Measure Status… Channel Lock… MFG Data Entering & Changing Variables Signal on mini-bargraphMeasure Back… Ppm O2 Measure Mode DisplayDisplay Status… Main… Channel BasicCal Module Manufacturing Data Displays LCDReset Abort Startup & InitializationInitiate quick start Quick StartMeasure Status… Back… Basic CONTROLS, Setup and Status Power supply voltages… Measure Back…Analyzer Channel Status Display Status… Main… ChannelDisplay Status… Single Component DisplayMain… Channel BasicCal Select Status… Tags Off LCDReset Multi Component DisplayBasic Controls and Setup Range numberOperation Status Range numberDisplay Controls Brightness and ContrastDisplay controls… Brightness 70 %Expert Controls and Setup Analyzer and I/O, expert controls & setup…Analyzer module controls… Measure Channel Back…Calibration Procedure Gas Scale FactorGas scale factor… Air background gas concentrationSpan Calibration Span gas 90.0 ppm Measure Status… Back…Alarms Alarms…Concentration alarm setup… Measure Back… ClrAlm Alarm generation isReRanges Ranges…Begin / end of ranges… Range Begin PpmMeasurement display configuration… Measurement Display ConfigurationLinearization Concentration Measurement Gas measurement unitsConcentration measurement… Gas concentration unitsTemperature compensation at span gas Sensor modelClamp low PVA to Ppm to mg/Nm3 conversion factorFlow measurement… Flow MeasurementFlow lower limit 400 ml/min Temperature measurement… Measure Status… Back… Temperature MeasurementTemperature measurement units System & Network I/O Module CON Trols Setup System & network I/O module controls…37.50 ppm System SIO module…Analog output setup… System SIOAnalog Output Setup Choose signalSelect analog output number Choose the analyzer moduleHold output during calibration Signal value for 0% outputHold output during calibration Outputs value on analyzer failure Operation modeOutput number Analog Output Updates Per Second Type of installed serial inter- faceOutput #1 Yes Baud rate 19200Relay Outputs Setup Invert signalCommunication protocol Choose source moduleChoose signal… System DIOFunction control Input numberSystem Configuration and DIAG- Nostics Load/Save Module Configura- tion Diagnostic MenusAnalyzer Module Diagnostics Date and Time Security CodesSystem reset Measure Back… System ResetEntering a level locked by security code Model NGA 2000 TO2 Overview Electrical Shock Hazard Section Maintenance and ServiceSensor Caustic Liquid Sensor Assembly Refilling SensorSensor Replacement Electrolyte ReplacementSensor Lleak Check Analyzer and I/O expert controls & setup… LinearizationDo not tilt Sensor excessively as elec- trolyte may leak Time & DateLinearization temperature values… Linearization…0000 ppm Linearization raw values sensor current…Flow Sensor Replacement Electronic ComponentsFuses Printed Circuit BoardsSection Troubleshooting TroubleshootingFailure to purge down to ppm levels Erratic and very insensitive readingsModel NGA 2000 TO2 Matrix Section Replacement PartsTO2 NGA 2000 TO2 Trace Oxygen Analyzer Module Replacement Parts Customer Service Section Return of MaterialReturn of Material TrainingModel NGA 2000 TO2 Section Index Model NGA 2000 TO2 Product ELECTROLYTE1 Emergency & first aid procedures Precautions to be taken in handlingOther precautions Threshold limit value Routes of entryRosemount Warranty Asia Pacific EUROPE, Middle EAST, Africa Latin America