Model NGA 2000 TO2
Instruction Manual
TO2 | 37.50 ppm |
|
Calibration…
Alarms…
Ranges…
Linearization…
Measurement display configuration…
Concentration measurement…
Flow measurement…
Temperature measurement…
Measure |
| Status… |
| Back… |
|
TO2
Y1:
Y2:
Y3:
Y4:
Measure
37.50 ppm
0.0000 ppm
1.050 ppm
7.000 ppm
72.800 ppm
Back…
Select Linearization…
TO2 | 37.50 ppm |
|
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure |
| Back… |
|
Select Linearization raw values (sensor cur- rent)…
TO2 | 37.50 ppm |
| |
|
|
X1: | 0.3828 uA |
X2: | 3.0400 uA |
X3: | 20.9300 uA |
X4: | 193.000 uA |
Measure |
| Back… |
|
Enter the data for each raw value point.
Return to the previous menu (Back…)
TO2 | 37.50 ppm |
|
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure |
| Back… |
|
Select Linearization set point values (ppm values)…
Enter the data for each set point value. Return to the previous menu (Back…)
TO2 | 37.50 ppm |
|
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure |
| Back… |
|
Select Linearization temperature values…
TO2 |
|
| 37.50 ppm |
| |
|
|
| |||
|
|
|
|
|
|
| T1: |
|
| 25.0 °C |
|
| T2: |
|
| 25.0 °C |
|
| T3: |
|
| 25.0 °C |
|
| T4: |
|
| 25.0 °C |
|
|
|
|
|
|
|
| Measure |
| Back… |
|
|
Enter the data for each temperature value point..
Return the the previous menu (Back…)
TO2 | 37.50 ppm |
|
Linearization raw values (sensor current)…
Linearization set point values (ppm values)…
Linearization temperature values…
Load linearization data!
Measure |
| Back… |
|
Select Load linearization data! press enter. The new data is now loaded into the analyzer and accurate measurements can be obtained.
Rosemount Analytical Inc. A Division of Emerson Process Management | Maintenance and Service |