Emerson NGA 2000 TO2 instruction manual Ranges…, Begin / end of ranges…, Range Begin Ppm

Page 42

Instruction Manual

760008-A September 2001

c. Ranges

Model NGA 2000 TO2

From the Analyzer Module Setup and Controls menu, choose Ranges as fol- lows:

Main Menu

Analyzer and I/O, expert controls and setup…

TO2

37.50 ppm

 

-- Analyzer Module Setup and Controls --

Calibration…

Alarms…

Ranges…

Linearization…

Measurement display configuration…

Concentration measurement…

Flow measurement…

Temperature measurement…

Measure

Status…

Back…

 

TO2

 

37.50 ppm

 

 

 

-- Ranges --

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Begin / end of ranges…

 

 

 

 

 

 

Actual range number:

3

 

 

 

Range and calibration control:

 

Local

 

 

 

 

 

 

 

Actual begin of range:

 

0.0 ppm

 

 

Actual end of range:

 

50.0 ppm

 

 

Measure

 

Back…

 

 

 

 

 

 

 

 

 

TO2

 

 

37.50 ppm

 

 

 

-- Begin / End of Ranges --

 

 

 

 

 

 

 

 

Range –

1 begin:

 

0.0 ppm

 

 

 

Range –

1 end:

 

10.0 ppm

 

 

Range –

2 begin:

 

0.0 ppm

 

 

Range –

2 end:

 

25.0 ppm

 

 

Range –

3 begin:

 

0.0 ppm

 

 

Range –

3 end:

 

50.0 ppm

 

 

Range –

4 begin:

 

0.0 ppm

 

 

Range –

4 end:

 

100.0 ppm

 

 

Measure

 

Back…

 

 

Move the cursor to the “Ranges…” line and press the key. The “Ranges” menu will dis- play as shown below.

The “Actual range number:” line on the Ranges menu shows the current range num- ber being used and allows changing of the range. The range values display below the line.

The “Range and calibration control:” function establishes the authority for setting the ana- lyzer range and starting calibration. The choices are:

Local: The range and calibration are con- trolled from the local platform. This is the normal mode of operation using the menus.

Program I/O module: The range and cali- bration are controlled by the programmable I/O board.

Inputs I/O module: The range and calibra- tion are controlled by the DIO digital inputs.

Move the cursor to the Begin/end of ranges..” line and press the key to reach the “Begin / End of Ranges” menu as shown.

Select any desired line and press the key to change any of the values using the and

keys or the and keys to select a digit.

Figure 3-19. Ranges

3-18 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Image 42
Contents Model NGA 2000 TO2 Essential Instructions Table of Contents Model NGA 2000 TO2System & Network I/O Module Controls Setup Material Safety Data SheetList of Illustrations Analyzer Module Interconnection with Instrument PlatformModel NGA 2000 TO2 Preface Intended USE Statement DefinitionsSafety Summary Model NGA 2000 TO2 Documentation CompliancesGlossary of Terms Softkeys SystemOverview Section Description and SpecificationsFeatures Typical ApplicationsTrace Oxygen Detector Coulometric Principle Trace Oxygen Analyzer Module Top View Sample SPECIFICATIONS1General PhysicalGas Connections Model NGA 2000 TO2 Electrolyte Section InstallationUnpacking Electrolyte AdditionGAS Overpressure GasesGAS Connections Sample FlowLeak Test Sensor DamageElectrical Connections Electrical Shock Hazard Analyzer Module Interconnection with Instrument Platform Outline and Mounting Dimensions Back Panel Connections Operation Displays & Operating KeysAnalyzer basic controls calibration & setup… Menu Lines & Softkey FunctionalityMeasure Status… Channel Lock… MFG Data Signal on mini-bargraph Entering & Changing VariablesMeasure Back… Measure Mode Display Ppm O2Display Status… Main… Channel BasicCal Module Manufacturing Data Displays LCDReset Abort Startup & InitializationQuick Start Initiate quick startMeasure Status… Back… Analyzer Channel Status Basic CONTROLS, Setup and StatusPower supply voltages… Measure Back… Display Status… Main… ChannelSingle Component Display Display Status…Main… Channel BasicCal Select Status… Tags Off LCDReset Multi Component DisplayOperation Status Basic Controls and SetupRange number Range numberDisplay controls… Display ControlsBrightness and Contrast Brightness 70 %Analyzer module controls… Expert Controls and SetupAnalyzer and I/O, expert controls & setup… Measure Channel Back…Calibration Procedure Gas Scale FactorGas scale factor… Air background gas concentrationSpan Calibration Span gas 90.0 ppm Measure Status… Back…Concentration alarm setup… Measure Back… ClrAlm AlarmsAlarms… Alarm generation isReBegin / end of ranges… RangesRanges… Range Begin PpmMeasurement Display Configuration Measurement display configuration…Linearization Concentration measurement… Concentration MeasurementGas measurement units Gas concentration unitsClamp low PVA to Temperature compensation at span gasSensor model Ppm to mg/Nm3 conversion factorFlow Measurement Flow measurement…Flow lower limit 400 ml/min Temperature Measurement Temperature measurement… Measure Status… Back…Temperature measurement units 37.50 ppm System & Network I/O Module CON Trols SetupSystem & network I/O module controls… System SIO module…Analog output setup… System SIOSelect analog output number Analog Output SetupChoose signal Choose the analyzer moduleSignal value for 0% output Hold output during calibrationHold output during calibration Operation mode Outputs value on analyzer failureOutput number Output #1 Yes Analog Output Updates Per SecondType of installed serial inter- face Baud rate 19200Communication protocol Relay Outputs SetupInvert signal Choose source moduleFunction control Choose signal…System DIO Input numberSystem Configuration and DIAG- Nostics Diagnostic Menus Load/Save Module Configura- tionAnalyzer Module Diagnostics Date and Time Security CodesSystem Reset System reset Measure Back…Entering a level locked by security code Model NGA 2000 TO2 Section Maintenance and Service Overview Electrical Shock HazardSensor Caustic Liquid Sensor Assembly Refilling SensorElectrolyte Replacement Sensor ReplacementSensor Lleak Check Do not tilt Sensor excessively as elec- trolyte may leak Analyzer and I/O expert controls & setup…Linearization Time & Date0000 ppm Linearization temperature values…Linearization… Linearization raw values sensor current…Fuses Flow Sensor ReplacementElectronic Components Printed Circuit BoardsFailure to purge down to ppm levels Section TroubleshootingTroubleshooting Erratic and very insensitive readingsModel NGA 2000 TO2 Section Replacement Parts MatrixTO2 NGA 2000 TO2 Trace Oxygen Analyzer Module Replacement Parts Return of Material Customer ServiceSection Return of Material TrainingModel NGA 2000 TO2 Section Index Model NGA 2000 TO2 Product ELECTROLYTE1 Other precautions Emergency & first aid proceduresPrecautions to be taken in handling Threshold limit value Routes of entryRosemount Warranty Asia Pacific EUROPE, Middle EAST, Africa Latin America