HoMedics BPA-110 manual Troubleshooting, Specifications

Page 15

TROUBLESHOOTING

If any abnormality arises during use, please check the following points

SYMPTOMS

POSSIBLE

CORRECTION

CAUSES

 

 

 

 

 

Unit does not turn on

Batteries have

Replace them with

when the

run down.

four new AA

START/STOP

 

alkaline batteries.

button is pushed.

 

 

Battery polarities

Re-insert the batteries

 

 

have been positioned

in the correct positions.

 

incorrectly.

 

 

 

 

EE measurement error

Cuff has been placed

Wrap the cuff properly

symbol shown on dis-

incorrectly.

so that it is positioned

play or the blood pres-

 

correctly.

sure value is displayed

 

 

excessively low

 

 

Did you talk or move

Measure again. Keep

(or high).

during measurement?

arm steady during

 

 

 

measurement.

 

Shaking of the arm with

 

the cuff on.

 

 

 

 

E1 error symbol shown

Air circuit abnormality.

Check cuff connection.

on display

Cuff tube may not be

Measure again.

 

plugged into monitor

 

 

correctly.

 

 

 

 

E2 error symbol shown

Inflation pressure

Switch the unit off, then

on display

exceeding 300 mmHg.

measure again.

 

 

 

E3 error symbol shown

Error determining

Measure again.

on display

measurement data.

 

 

 

 

Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations Department can be found on the warranty page.

SPECIFICATIONS

Power Source:

Four 1.5V DC (AA) batteries or 6V DC AC adapter

 

 

Measurement Method:

Oscillometric

 

 

Measurement Range:

Pressure: 40~280 millimeters mercury (mmHg)

Pulse: 40~199 beats/minute

 

 

 

Accuracy:

Pressure: ±3 mmHg

Pulse: ±5% of reading

 

 

 

Pressure Sensor:

Semi-conductor

 

 

Inflation:

Pump driven

 

 

Deflation:

Automatic Pressure release valve

 

 

Memory Capacity:

60 memories for each User (120 total)

 

 

Auto-shut-off:

1 minute after last button operation

 

 

Operation

Temperature: 50°F~104°F (10°C ~40°C)

Environment:

Humidity: Less Than 85% RH

 

 

Storage Environment:

Temperature: -4°F~140°F (-20°C~60°C)

 

Humidity: Less Than 85% RH

Weight:

0.72 lb (325 g) (without batteries)

 

 

Arm Circumference:

Standard size cuff: 9"–13" (23-33 cm)

Large size cuff:13"–17" (33-43 cm)

 

 

 

Dimensions:

5.9"(L) x 4.4"(W) x 2.2"(H)

149mm(L) x 111.5mm(W) x 57mm(H)

 

 

 

 

Internally Powered equipment

 

Type BF Equipment

 

IPx0-No special protection against the ingress of

 

water

 

Not suitable for use in presence of flammable

 

anesthetic mixture with air, oxygen or nitrous

 

oxide.

 

Continuous operation with short-time loading.

 

 

Note: These specifications are subject to change without notice.

27

28

Image 15
Contents Deluxe Automatic Blood Pressure Monitor Table of Contents Important Product Notices and Safety Instructions About Blood Pressure NAME/FUNCTION of Each Part HOW this Blood Pressure Monitor WorksDisplay Explanations Display SymbolsInstalling Batteries Using the AC AdapterReplace the batteries if Measurement Date & Time SET Procedure Using the ARM CuffPlug the cuff tube into the left side of the unit Important Notes Measurement ProcedureMeasurement is then automatically stored into memory Risk Category IndicatorRecalling Values from Memory Clearing Values from Memory Press and hold the Date & Time set buttonsImportant Notes Regarding Your Blood Pressure Measurement Potential for Electromagnetic Interference CARE, Maintenance & CleaningSpecifications TroubleshootingMonitor de lujo Automático de Presión Sanguínea Información DE USO Información InicialAvisos Importantes DEL Producto E Instrucciones DE Seguridad Sobre LA Presión Sanguínea NOMBRE/FUNCIÓN DE Cada Pieza Cómo Funciona Este Monitor DE Presión SanguíneaSímbolos de la pantalla Explicaciones DE LA PantallaUSO DEL Adaptador DE CA Instalación DE LAS PilasUSO DEL Brazalete Procedimiento DE Configuración DE Fecha Y HoraNota Notas importantes Procedimiento DE MediciónLa medida se almacenará automáticamente en la memoria Nota Indicador DE Catergoria DE RiesgoCómo Recuperar LOS Valores DE LA Memoria Cómo Borrar LOS Valores DE LA Memoria Potencial DE Interferencia Electromagnética CUIDADO, Mantenimiento Y LimpiezaEspecificaciones Diagnóstico Y Resolución DE Problemas