A&D TM-2655P manual Specifications, Performance specifications

Page 13

1-3 Specifications

Table 1 Specifications

Performance specifications

 

 

General

 

 

 

 

 

 

 

 

 

 

 

Power supply

120VAC

60 Hz

(For USA area)

 

 

 

230VAC

50 Hz

(For Europe area)

 

 

 

(factory-preset)

 

 

 

 

 

 

 

 

Power consumption

Approx. 40VA

 

 

 

 

 

 

 

 

Protection against electrical shock

Class I , Type B

 

 

 

 

 

 

 

 

 

Display

LED

 

 

 

 

 

 

 

 

 

EMC

EN 60601-1-2

 

 

 

 

 

 

 

 

 

Blood pressure measurement

 

 

 

 

 

 

 

 

 

Blood pressure measuring method

Oscillometric method

 

 

 

 

 

 

Pressure detection method

Capacitance type pressure transducer

 

 

 

 

 

 

 

Pressure range

0 – 300 mmHg

 

 

 

 

 

 

 

Accuracy

Pressure: ±3 mmHg

 

 

 

Pulse rate: ±5%

 

 

 

 

 

 

 

Measurement range

Blood pressure: 10 – 280 mmHg

 

 

 

Pulse rate: 30 – 200 bpm

 

 

 

 

 

 

 

Pressurizing method

Micro pump

 

 

 

 

 

 

 

Air pressure control method

Rubber valve, ceramic valve

 

 

 

 

 

 

Rapid air exhaust system

Electromagnetic valve

 

 

 

 

 

 

Safety mechanism

The electromagnetic valve is released when

 

 

 

approx. 320 mmHg or greater is detected.

 

 

 

 

 

 

Communications functions

Serial output: RS232C level

 

 

 

 

 

 

Environment specifications

 

 

 

 

 

 

 

 

 

Operating temperature and humidity

10 to 40°C, 85%RH or less, non condensing

 

 

 

 

 

 

Storage temperature and humidity

-20 to 60°C, 95%RH or less, non condensing

 

 

 

 

 

 

Physical specifications

 

 

 

 

 

 

 

 

 

Overall dimensions

245(W) X 320(H) X 390(D) mm

 

 

 

 

 

 

 

Weight

Approx. 9 kg

 

 

 

 

 

 

 

5

Image 13
Contents Automatic Blood Pressure Monitor Page Important information that helps users operate the device Compliance with the Australian EMC Frame Work ComplianceGrounding Safety PrecautionsWhen installing and storing the device General PrecautionsMust be disposed of according to the applicable regulations Page Contents Troubleshooting External Dimensions Introduction FeaturesItems contained Unpacking and InspectionPerformance specifications SpecificationsMain Unit Part NamesRear Panel SymbolsProper installation InstallationBe sure to insert your arm to the shoulder OperationMeasurement results Clock Adjustment Mode Setting the ClockSetting Example Page Paper installation procedure Installing the Printer PaperFunction Setting Mode Setting the FunctionsF02 Off Description of the FunctionsLine printing Table printing Bit pattern printing Printing SamplesChannel 1 Miniature DIN 8-pin Communication SpecificationsChannel 2 D-sub 9-pin Cable connection diagramReplacing the Arm Cuff Cover MaintenanceReplacing the Fuses Error CodesChecking the Counter Cleaning MaintenanceOptions and Consumables Troubleshooting External dimensions External Dimensions